background image

CALIBRATION

Equipment required:  3-1/2 or 4-1/2 digit multimeter with

±0.1% accuracy, stable voltage source, TRC III Ice Point

TM

Cell, TRP (J, K, T) Reference Probe. 
Set up the equipment as shown below.  Set the Function

Switch to “°C” or “°F” position.  Turn power on.

Set voltage source and adjust potentiometers as follows. 

Set Voltage Source Adjust for Multimeter Reading

Model TAC80B-J

°C

0.000 mV

P1

00.00 mV

33.102 mV

P2

600.0 mV

°F

0.000 mV

P4

32.00 mV

33.102 mV

P3

1112.0 mV

Model TAC80B-K

°C 

0.000 mV

P1

00.00 mV

37.325 mV

P2

900.0 mV

°F

0.000 mV

P4

32.00 mV

37.325 mV

P3

1652.0 mV

Model TAC80B-T

°C

0.000 mV

P1

00.00 mV

14.862 mV

P2

300.0 mV

°F

0.000 mV

P4

32.0 mV

14.862 mV

P3

572.0 mV

Figure 1-2.  Potentiometer Locations

DIGITAL MULTIMETER

V

COM

A

COM

+

INPUT

K TYPE

THERMOCOUPLE LEADS

K TYPE

THERMOCOUPLE CONNECTOR

TRC III

ICE POINT GENERATOR

SERIES TRP REFERENCE PROBES

COPPER CONNECTOR

COPPER LEADS

100mV

10mV

1mV

100

µ

V

10

µ

V

STABLE VOLTAGE SOURCE

1

µ

V

0

0

4

0

9

5

V

Figure 1-3.  Calibration Set-up

Servicing North America:

U.S.A.:

Omega Engineering, Inc., One Omega Drive, P.O. Box 4047

ISO 9001 Certified

Stamford, CT 06907-0047

Tel: (203) 359-1660

FAX: (203) 359-7700

e-mail: [email protected]

Canada:

976 Bergar

Laval (Quebec), Canada H7L 5A1

Toll-Free: 1-800-826-6342

TEL: (514) 856-6928

FAX: (514) 856-6886

e-mail: [email protected]

For immediate technical or application assistance:

U.S.A. and Canada:

Sales Service: 1-800-826-6342/1-800-TC-OMEGA

®

Customer Service: 1-800-622-2378/1-800-622-BEST

®

Engineering Service: 1-800-872-9436/1-800-USA-WHEN

®

Mexico:

En Español: 001 (203) 359-7803

FAX: (001) 203-359-7807

[email protected]

e-mail: [email protected]

Servicing Europe:

Benelux:

Managed by the United Kingdom Office

Toll-Free: 0800 099 3344

TEL: +31 20 347 21 21

FAX: +31 20 643 46 43

e-mail: [email protected]

Czech Republic:

Frystatska 184

733 01 Karviná, Czech Republic

Toll-Free: 0800-1-66342

TEL: +420-59-6311899

FAX: +420-59-6311114

e-mail: [email protected]

France:

Managed by the United Kingdom Office

Toll-Free: 0800 466 342

TEL: +33 (0) 161 37 29 00

FAX: +33 (0) 130 57 54 27

e-mail: [email protected]

Germany/Austria:

Daimlerstrasse 26

D-75392 Deckenpfronn, Germany

Toll-Free: 0 800 6397678

TEL: +49 (0) 7059 9398-0

FAX: +49 (0) 7056 9398-29

e-mail: [email protected]

United Kingdom:

OMEGA Engineering Ltd.

ISO 9001 Certified

One Omega Drive, River Bend Technology Centre, Northbank 

Irlam, Manchester M44 5BD England

Toll-Free: 0800-488-488

TEL: +44 (0)161 777-6611

FAX: +44 (0)161 777-6622

e-mail: [email protected]

OMEGAnet

®

On-Line Service 

Internet e-mail

omega.com 

[email protected]

FOR 

WARRANTY

RETURNS, please have the following

information available BEFORE contacting OMEGA:

1. Purchase Order number under which the product was

PURCHASED,

2. Model and serial number of the product under warranty, and

3. Repair instructions and/or specific problems relative to the

product.

FOR 

NON-WARRANTY

REPAIRS, consult OMEGA for current repair

charges. Have the following information available BEFORE

contacting OMEGA:

1.  Purchase Order number to cover the COST of the repair,

2. Model and serial number of the product, and

3. Repair instructions and/or specific problems relative to the

product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the lat-

est in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2009 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced

to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.

WARRANTY/ DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 

13 months

from

date  of  purchase.  OMEGA’s  WARRANTY  adds  an  additional  one  (1)  month  grace  period  to  the  normal 

one  (1)  year  product

warranty

to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. 

If  the  unit  malfunctions,  it  must  be  returned  to  the  factory  for  evaluation.  OMEGA’s  Customer  Service  Department  will  issue  an

Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be

defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the

purchaser,  including  but  not  limited  to  mishandling,  improper  interfacing,  operation  outside  of  design  limits,  improper  repair,  or

unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of

having  been  damaged  as  a  result  of  excessive  corrosion;  or  current,  heat,  moisture  or  vibration;  improper  specification;

misapplication;  misuse  or  other  operating  conditions  outside  of  OMEGA’s  control.  Components  in  which  wear  is  not  warranted,

include but are not limited to contact points, fuses, and triacs.

OMEGA  is  pleased  to  offer  suggestions  on  the  use  of  its  various  products.  However,  OMEGA  neither  assumes

responsibility  for  any  omissions  or  errors  nor  assumes  liability  for  any  damages  that  result  from  the  use  of  its

products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the

parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER  WARRANTIES

OR  REPRESENTATIONS  OF  ANY  KIND  WHATSOEVER,  EXPRESSED  OR  IMPLIED,  EXCEPT  THAT  OF  TITLE,  AND  ALL

IMPLIED  WARRANTIES  INCLUDING  ANY  WARRANTY  OF  MERCHANTABILITY  AND  FITNESS  FOR  A  PARTICULAR

PURPOSE  ARE  HEREBY  DISCLAIMED.  LIMITATION  OF  LIABILITY:  The  remedies  of  purchaser  set  forth  herein  are

exclusive,  and  the  total  liability  of  OMEGA  with  respect  to  this  order,  whether  based  on  contract,  warranty,

negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon

which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR

21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s)

be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes

no  responsibility  as  set  forth  in  our  basic  WARRANTY/ DISCLAIMER  language,  and,  additionally,  purchaser  will  indemnify  OMEGA

and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS / INQUIRIES

Direct  all  warranty  and  rep

air  requests/inquiries  to  the  OMEGA  Customer  Service  Department.  BEFORE  RETURNING  ANY

PRODUCT(S)  TO  OMEGA,  PURCHASER  MUST  OBTAIN  AN  AUTHORIZED  RETURN  (AR)  NUMBER  FROM  OMEGA’S  CUSTOMER

SERVICE  DEPARTMENT  (IN  ORDER  TO  AVOID  PROCESSING  DELAYS).  The  assigned  AR  number  should  then  be  marked  on  the

outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit. 

It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI 

regulations that apply. OMEGA is constantly pursuing certification of its products to the 

European New Approach Directives. OMEGA will add the mark to every appropriate device 

upon certification.
The information contained in this document is believed to be correct, but OMEGA 

Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to 

alter specifications without notice. 

WARNING:

 

These products are not designed for use in, 

and should not be used for, human applications.

Отзывы: