18 |
P a g e
5.1.7.1
Sensor Trigger Function
The Sensor Trigger function is used to reset the IPSO min/max values as well as controlling the Calibration
process.
Sensor Trigger Function
7
6
5
4
3
2
1
0
0
0
0
0
0
0
0
Reset
Min/Max
15
14
13
12
11
10
9
8
0
0
Calibration
Reset
Calibration
Status
Calibration
Mode
Capture
High
Capture
Low
Calibration
Start
Setting the Reset Min/Max bit to 1 will reset the Min/Max values recorded by the IPSO process.
The Calibration mode is entered by writing a 1 to the Calibration Mode bit. While in the calibration mode the
calibration registers may be accessed and the Calibration Start may be set. When the Calibration Start bit is set
the Calibration Status bit will remain set until the calibration process is complete.
Setting the Calibration Reset bit will clear the calculated Offset value.
5.2
Thermocouple Descriptor
The Thermocouple Input interface provides interfaces to type J, K, T, E, N, R, S, B, and C thermocouples. The default T type is
typically used on Heat Flux sensor devices (HFS-05). The Sensor Configuration and Sensor Device fields may be written to
provide control of the overall function of the channel and the thermocouple types used.
Offset
Name
Value
Description
0x00
Measurement
Type
0x20
Temperature (°C)
0x01
Data Type/Format
0x26
Float,
Factory Calibration enabled
0x02
Configuration
0x4?
Determines Thermocouple type
0x03
Sensor Device
0x??
Determines connection type
0x04..0x08
UOMR
“°C”
Units of measure
5.2.1
Thermocouple Measurement Types
The Thermocouple interface provides a measurement of Temperature in °C.
Sensor
Type
SI Derived
Units
Measurement
0x20
°C
Temperature