5.1.2.2
Factory Calibrate
The Factory Calibrate bit is used during factory calibration. If set, the factory calibration attributes are applied to the
sensor reading before the sensor scaling. If clear, not factory calibration attributes are applied. If the sensor does not
support factory calibration the Factory Calibrate bit is ignored and will always read as 0.
5.1.2.3
Sensor Writeable
If the Sensor Writeable bit is set the sensor value may be overwritten with a preset value. This capability is useful in
sensors such as up/down counters, where a preset, or possibly a zero value must be written to the sensor value.
5.1.2.4
Smart Sensor
Refer to the
Smart Sensor Device Interface
documentation.
5.1.3
Load Cell Configuration
Load Cell Configuration
7
6
5
4
3
2
1
0
Available
Assigned
Apply
Scaling
Lock
Sensor Type (Range)
(Not Used)
0
*
?
?
5.1.3.1
Apply Scaling
If set, the user defined Offset and Gain values will be used to adjust the sensor reading:
Result = (Raw Reading * Gain) + Offset
5.1.3.2
Lock
If set, the user specified units of measure string (4-character maximum) will be used in place of the default units of
measure.
5.1.3.3
Assigned
Refer to the
Smart Sensor Device Interface
documentation.
5.1.3.4
Available
Refer to the
Smart Sensor Device Interface
documentation.
5.1.4
Load Cell Device Byte
The Sensor Device byte determines whether the DIO TARE function is enabled for devices with DIO available. If the bit
is enabled than the corresponding DIO signal will act as a remote TARE trigger.
The Open Detect bit allows enabling
the open circuit detection function.
Load Cell Device Byte
7
6
5
4
3
2
1
0
0
0
0
Open Circuit Detect
0
Disabled
1
Enabled
DIO TARE ENABLE
0
0
DIO_1
DIO_0