3. To end automatic data logging
, first press the
LOGGER
button to pause data logging.
Then press the
REC/ENTER
button and hold it for at least two seconds. This will cause
“REC” to disappear.
4. To log data manually
, set the sampling time to zero using Step 9 of the Setup
Instructions on p. 8. Then press the
REC/ENTER
button once. The display will show the
term “REC” on the top line and a flashing letter “P” and a number from 1 to 99 on the
bottom line. The number indicates the position on the SD card that will be used to store
manually logged data. In manual data logging mode, this position can be changed.
5. To change the storage location of manually logged data
, press the
TYPE
button
once; this will cause the “P” to disappear from the left side of the display and the value
on its line to begin flashing. Once the flashing begins, you can use the
TYPE
and
PAGE
buttons to change the flashing value to any number between 1 and 99. Once
you have chosen the storage location, press the
REC/ENTER
button to save the setting.
This will cause the value to stop flashing and the “P” to return.
6. To end manual data logging
, press the
REC/ENTER
button and hold it for at least three
seconds. This action will cause “REC” to disappear from the top line of the display.
7. To execute the loop data recording function
(if the function was enabled during setup)
press the
REC/ENTER
button while the instrument is in measurement mode. Doing so
will make the term “REC” appear at the top of the display and prepare the instrument to
log data during the specified daily period.
8. To pause loop data recording
during the specified period, press the
LOGGER
button
once. This will make the term “REC” stop flashing. Pressing the
LOGGER
button again
will resume loop data recording.
9. To end loop data recording
, press the
REC/ENTER
button and hold it for at least
three seconds. This will cause “REC” to disappear from the display and end loop data
recording.