Omega PRESSURE GAUGES DPG5000 Скачать руководство пользователя страница 6

• Instructions

NOTE:.All.units.are.factory.calibrated.prior.to.shipment.

1. Zero Trimming

NOTE:.Some.units.are.equipped.with.zero.“knob”.

If it becomes necessary to re-adjust the “Zero” on the display, this can be 

accomplished by turning the trimpot marked “Z” just to the left of the On/Off 
push button. On the DPG5000 there is either a zero adjustment knob or a black 
nylon cover screw. If the gauge has a cover screw, it is necessary to remove the 
cover screw to access the zero adjustment pot. An ideal zero is indicated by a 
reading of 000 with an intermittently flashing “-” sign.

NOTE:.A.jewelers.screwdriver.or.an.eyeglass.repair.screwdriver.(supplied).

is.a.suitable.instrument..Do.not.make.changes.to.the.Span.adjustment.(the.“S”.
pot.to.the.right.of.the.push.button).as.part.of.the.zero.trimming..The.Span.should.
only.be.changed.as.part.of.the.re-calibration.of.a.gauge.with.a.known.pressure.
source.

2. Use of Auto-Tare Feature

When the DPG5000 is first turned on you will notice that the display flashes 

for the first 4 seconds. If the On/Off button is pushed and held during this 
period, any existing reading will be re-set to zero. This allows the user to “Tare-
Out” a starting weight in hydraulic weighing applications. To remove the Tare 
feature turn the unit off and back on again and do not push the On/Off button 
during the flashing period.

3. Battery Replacement (9 Volt Type)

The battery can be replaced simply by removing the single screw at the top of 

the battery door. Remove the old battery, unplug the cable connector and replace 
with Eveready type 216 or equivalent. Replace the battery door and secure with 
the self-tapping screw. (Do not overtighten)

NOTE:.For.best.accuracy,.re-calibration.per.step.4.should.be.performed.at.the.

time.batteries.are.replaced.

 

DPG5000 

M3366/0606

Содержание PRESSURE GAUGES DPG5000

Страница 1: ...DPG5000 PRESSURE GAUGES M3366 0606 User s Guide http www omega com e mail info omega com...

Страница 2: ...ine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Eq...

Страница 3: ...mming 4 Use of Auto Tare Feature 4 Battery Replacement 4 Re Calibration 5 Use Of Peak Hold Feature Optional 5 Analog Output Option 0 To 2 VDC 5 4 20 mA Transmitter Option 6 External Power 24 VDC and 0...

Страница 4: ...GA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA either ver...

Страница 5: ...kaging to prevent breakage in transit Warranty Returns For WARRANTY RETURNS please have the following information available BEFORE contacting OMEGA 1 Purchase Order number under which the product was...

Страница 6: ...he Span should only be changed as part of the re calibration of a gauge with a known pressure source 2 Use of Auto Tare Feature When the DPG5000 is first turned on you will notice that the display fla...

Страница 7: ...hold feature and causes an arrow to appear in the display In this mode only the highest reading sensed will be shown until a higher reading comes along This peak reading will be retained until the uni...

Страница 8: ...rcuit protection Power supply voltage must be sufficient to maintain a minimum of 9 VDC at the gauge terminals after dropping voltage across RL at full scale current 20 mA Example If RL 250 ohm then d...

Страница 9: ...T PIN D MINUS 24 VDC POWER IN PIN E NO CONNECTION PIN F CASE GROUND 9 External Power Without Other Options The external power input is accessed through a miniature Phone Jack mate supplied on the back...

Страница 10: ...DPG5000 M3366 0606 Notes...

Страница 11: ...irectives OMEGA will add the CE mark to every appropriate device upon certification The information contained in this document is believed to be correct but OMEGA Engineering Inc accepts no liability...

Страница 12: ...44 e mail nl omega com Czech Republic Rud arm dy 1868 733 01 Karvin TEL 420 0 69 6311899 FAX 420 0 69 6311114 Toll Free in Czech Rep 0800 1 66342 e mail czech omega com France 9 rue Denis Papin 78190...

Отзывы: