Calibrate the Controller:
1.
If it’s not already shown, press
MENU
until the
controller displays
r
rd
d..c
cf
f
.
2.
Press
T/mV
. The controller
displays
r
r..1
1
=
2
2
or
r
r..1
1
=
3
3
(2 or 3 point calibration).
3.
Press
/pH
to select the
desired calibration type.
4.
Press
T/mV
to display
r
r..2
2
=
f
f
or
r
r..2
2
=
c
c
(temperature unit of F or C).
5.
Press
/pH
to select the
desired temperature unit.
6.
Press
MENU
to select the calibration. The unit
displays.
7.
Proceed with Calibration depending on
Temperature Compensation Setting:
•
If you choose OFF, go to “To Perform
Calibration” section.
•
If you choose METR, ensure the RTD is
properly connected and go to “To Perform
Calibration” section.
•
If you choose MANU, configure temperature
as described in the next section.
T/mV
MENU
RESET
/pH
SETPTS
T/mV
MENU
RESET
/pH
SETPTS
SETPTS
T/mV
MENU
RESET
/pH
=
Connect the pH Electrode
1.
Secure the pH electrode to the pH input BNC
connector on the back of the controller.
2.
If you are using automatic temperature
compensation (ATC), connect the RTD as shown
below.
Apply Power
1.
Apply ac power to the unit.
The unit initializes, flashing the following
messages:
r
rs
st
t
,
p
ph
h
,
i
in
ni
it
t
. Then a pH value
appears.
2.
Verify that a value appears. If not:
•
Remove ac power.
•
Verify the TB1 power connections.
•
Check your power source.
•
Apply ac power again.
TB1
TB2
1 2
3 4 5 6
7 8
9 10
1 2
3
11 12
RTD
+ -
TB1
TB2
1 2
3 4 5 6
7 8
9 10
1 2
3
11 12
+ -
pH Input
pH Electrode
START HERE
2
3
4
Using This Quick Start Manua
l
Use this Quick Start manual to set up your pH
Controller and begin operation. Information is provided
on how to:
•
Mount the controller
•
Connect ac Power
•
Connect a pH electrode
•
Calibrate the controller prior to use
For complete information on this controller, refer to the
Operator’s Manual.
Before You Begin
In addition to the meter and the related parts, you will
need the following items to set up your meter:
•
ac power, as listed on meter’s ID/Power Label
•
pH electrode (with BNC input connector)
•
1
⁄
8
” flat blade screwdriver
Mount the Unit
1. Cut a panel opening using
the dimensions shown to
the right.
2. Position the unit in the
opening, making sure the
front bezel gasket is flush
to the panel.
3. Slide on mounting bracket to secure.
Connect
~
ac Power
Warning: Do not connect AC power to your device until you
have completed all input and output connections. This
device must only be installed by a specially trained
electrician with corresponding qualifications. Failure to
follow all instructions and warnings may result in injury!
1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the
following figure and tighten the lockdown screws.
4. Tug gently on the wires to verify the connections.
To Select Temperature
Compensation Mode:
1.
Press
MENU
. The controller
displays:
2.
Press
T/mV
to display the
current setting.
3.
Press
/pH
to change to the
desired setting. Choices are:
OFF = the controller uses a
constant 25ºC for temperature
compensation.
MANU = the controller uses a manually entered
value for temperature compensation.
METR = the controller uses the RTD input for
automatic temperature compensation.
4.
Press
MENU
to select the temperature
compensation setting shown. The controller
displays.
To Set The Decimal Point Position:
1.
If it’s not already shown, press
MENU
until the controller
displays
d
de
ec
c..p
p
.
2.
Press
T/mV
. The controller
displays:
3.
Press
/pH
to move the decimal point to the
desired location. The choices are
f
ff
f..f
ff
f
,
f
ff
ff
f..f
f
, or
f
ff
ff
ff
f
.
4.
Press
MENU
to select the decimal point position
shown. The controller displays:
T/mV
MENU
RESET
/pH
SETPTS
T/mV
MENU
RESET
/pH
SETPTS
SETPTS
T/mV
MENU
RESET
/pH
T/mV
MENU
RESET
/pH
SETPTS
T/mV
MENU
RESET
/pH
SETPTS
T/mV
MENU
RESET
/pH
SETPTS
T/mV
MENU
RESET
/pH
SETPTS
T/mV
MENU
RESET
/pH
SETPTS
=
45,00 + 0,61/-0,00
(1.772 + .024/–.000)
92,00 + 0,81/–0,00
(3.622 + .032/–.000)
PANEL THICKNESS
1,5
R(.06)
4 PLCS
6,4 (.25) MAX
0,8 (.03) MIN
NOTE: Dimensions in Millimeters (Inches)
TB1
TB2
1 2
3 4 5 6
7 8
9 10
1 2 3
11 12
FUSE
EARTH GROUND
GREEN WIRE
AC Power
Check for proper Earth grounding in the power
distribution system (single phase).
SWITCH
LINE
NEUTRAL
EARTH
(L) LINE
(N) NEUTRAL
pH INPUT