7

MAINTENANCE

 

Our customer service representatives are available for application assistance, calibration, repair, and 
solutions to specific problems. Contact our Service Department before returning any equipment. In 
many cases, problems can be solved over the telephone. If the sensor is not performing as it should, 
try to match the symptom below to the problem. If the table does not help, call Omega for further 
advice.

Troubleshooting
Symptom

Probable Cause

Solution

Sensor cannot be 

found, or software will 

not start

Software or USB 

driver failed to install 

properly

Disconnect sensor. Uninstall software. 

Reinstall as Administrator (right-click 

the installer and select “Run As 

Administrator”). Do not connect the 

sensor until prompted by the installer.

Software will not start, 

or will not install

Incompatible 

Windows version

Try installing on a PC with Windows 

Vista, 7, 8, or 10

Erroneous 

temperature

Field of view 

obstruction

Remove obstruction

LENS CLEANING

Keep the lens clean at all times. Any foreign matter on the lens would affect measurement accuracy.  
Blow off loose particles (if not using the air purge accessory) with an air ‘puffer’.

OPERATION

Once the sensor is in position and the appropriate power, air, water, and cable connections are secure, 
the system is ready for continuous operation by completing the following simple steps:

1.  Turn on the power supply
2.  Turn on the Modbus Master
3.  Read / monitor the temperature

IMPORTANT

Be aware of the following when using the sensor:
•  If the sensor is exposed to significant changes in ambient temperature (hot to cold, or cold to hot), 

allow 20 minutes for the temperature to stabilise before taking or  recording measurements.

•  Do not operate the sensor near large electromagnetic fields (e.g. around arc welders or induction 

heaters). 

  Electromagnetic interference can cause measurement errors.
•  Wire must be connected only to the appropriate terminals.

Содержание OS-MINIUSB series

Страница 1: ...e mail info omega com For latest product manuals omegamanual info Shop online at omega comSM User sGuide OS MINIUSB SERIES...

Страница 2: ...notice Servicing North America U S A Omega Engineering Inc One Omega Drive P O Box 4047 Stamford CT 06907 0047 USA Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada o...

Страница 3: ...21 20 1 OS MINIUSB SN201 Interface USB Accuracy 1 of reading or 1 C whichever is greater Repeatability 0 5 of reading or 0 5 C whichever is greater Emissivity 0 2 to 1 0 Response Time t90 125 ms 90 re...

Страница 4: ...collar Laser sighting tool OPTIONS An optional certificate of calibration is available for newly manufactured sensors If required it must be ordered at the same time as the sensor OPTICAL CHART The op...

Страница 5: ...CE To minimise electromagnetic interference or noise the sensor should be mounted away from motors generators and such like MECHANICAL INSTALLATION All sensors come with a 1 5 m USB cable and a mounti...

Страница 6: ...ntil the software has been fully installed The software will tell you when to connect the sensor This will ensure the USB driver for the sensor is installed properly 1 Insert the provided CD or downlo...

Страница 7: ...ivity setting should match the actual emissivity of the target surface Non reflective non metals such as rubber foods thick plastics organic materials and painted surfaces generally have a high emissi...

Страница 8: ...Status bits active high Bit 0 Measurement error Bit 1 Sensor temperature low Bit 2 Sensor temperature high Bit 3 Object temperature low Bit 4 Object temperature high R 0x16 1 Reflected Temperature no...

Страница 9: ...with Windows Vista 7 8 or 10 Erroneous temperature Field of view obstruction Remove obstruction LENS CLEANING Keep the lens clean at all times Any foreign matter on the lens would affect measurement a...

Страница 10: ...8...

Страница 11: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Страница 12: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Отзывы: