Product Maintenance
Battery Replacement
Materials:
OM-CP-BAT110
1. Unscrew the bottom of the logger and remove the battery.
2. Place the new battery into the logger. Note the polarity of the battery.
3. Screw the cover back onto the logger.
Recalibration
The OM-CP-HITEMP140-FP standard calibrations are two points at 30°C and 140°C.
Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in
the software when the device is due.
Notice: Steam Sterilization Applications
The pervasive nature of pressurized steam creates a very difficult environment for electronics.
Please refer to the following preventative maintenance procedure when using this device in steam sterilization
applications.
Additionally, this device is not ideal for steam sterilization applications above 121 °C/1.1 bar.
PREVENTATIVE MAINTENANCE
After every 3 hours of steam exposure:
1. Remove the endcap and battery from the device (ref. battery change procedure on Product User Guide)
2. Place open logger (minus battery) in an oven at 120°C (250°F) for a minimum of 30 minutes
3. Remove logger from oven and allow to cool to room temp
4. Re-assemble the logger with the battery (note polarity) and endcap
Quick Start Manual
5