THERMOCOUPLE FIELD CALIBRATION SUBMENU:
RTD and Process are perfectly calibrated. This section is applicable to
Thermocouple (TC) calibration
only
.
Be sure that the TC being used to calibrate the meter is of the type
selected in the TC submenu. Place the TC in an ice-bath (or other
0°C / 32°F
environment). In ambient temperature conditions: connect the
TC to the meter, apply power to the meter.
CAUTION:
Do not proceed with TC calibration unless the above conditions
have been in effect for at least one hour.
Press
a
7)
Display shows
CAL1
.
Press
d
8)
Display shows flashing
0000
.
Press
a
*
9)
Display will still show flashing
0000
.
Press
d
*
10)
Display shows
OUT1
(meaning Calibration is complete)
*
If you accidently engage the flashing
0000
(CAL° alert) simply re-press
the last button you pressed, to avoid unintentionally mis-calibrating your
meter.
CAUTION:
Do not perform the following steps until you fully
understand this entire section.
°