CALIBRATION PROCEDURE
Note:
The following calibration procedures should be performed only by
qualified technicians who have access to the following:
Equipment: The class of calibration should be 10 times greater than the
measured meter.
Step 1: Short the J3 with jumper.
Step 2: Apply 65 mV to the T1 input from Volts Calibrator and wait till the
reading becomes stable, then press the SET key until the display shows 0.1.
Step 3: Apply Type K 0°C to the input T2 and wait till the reading
becomes stable, then press the SET key until the meter shuts off by itself.