Omega FSW-6000 Series Скачать руководство пользователя страница 9

®

Pre-Installation

Pre-Installation Checks:

1) Its recommended that the flow switch is
installed with a distance of ½ a meter of the pipe
bent where the flow enters and 5x times the
diameter of the pipe where the flow exits, enabling
it to have an accurate reading (Fig. 1).

Verify that the installation point isn

t near any

connections, valves, elbows or anything similar,
this can cause errors in the reading of the probe
due to turbulence in the pipe.

2) It is important that the flow switch is not installed
at the highest point in the pipe run or in a location
where there is the risk of air accumulating in the
pipe. Keep in mind that the ideal mounting
location is where the pipe is always full. This will
ensure that the switch is always immersed in the
flow. (Fig. 2 correct, fig. 3 incorrect)

3) In pipes that have pressure pumps or retention
valves, we recommend that the probe be installed
before the pump due to the fact that it will have
less turbulence. (Fig. 4)

4) Confirm that the wire connections are correct
and that the available power supply is compatible
with the FSW unit.

5) Verify that the operating pressure and
temperature of the process corresponds to the
operating parameters of the FSW unit.

More recommendations and handling instructions
can be found on page 14.

Fig. 1

Fig. 4

Pump

Fig. 2

Fig. 3

1/2m

5xØ

09

www.omega.com e-mail: [email protected]

Содержание FSW-6000 Series

Страница 1: ...m For latest product manuals www omegamanual info Shop online at omega comSM User sGuide FSW 6000 FSW 7000 Series Thermal Dispersion Flow Switches This datasheet has been downloaded from http www digc...

Страница 2: ...ervice 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide omega com info omega com The information contained in this d...

Страница 3: ...duction 4 Models Dimensions 5 Wiring Diagram 6 Pre Installation 9 FSW Relay Status Guide 8 Installation 10 Calibration 11 Handling 13 Technical Specifications 14 Trouble Shooting 16 www omega com e ma...

Страница 4: ...e switch point status of the unit The sensing element and connection of the FSW are made with 316 S S and can be coated when necessary The enclosure is offered in either nylon or aluminum All models c...

Страница 5: ...ocess Connections TC Connection ANSI 150 ANSI 300 Rubber Seal Process Connection NPT BSP FF RF 100mm 100mm 100mm 50mm 45mm 35mm 50mm 75mm 12mm 12mm 12mm 3 4 NPT 1 NPT 1 NPT NPT FSW w insertion length...

Страница 6: ...mmon Green NC Black Power Supply Contacts Ground FSW6000 AC Powered 2000mm 1 2 Connector 12mm NO or NC contact 2000mm 1 2 Connector 12mm _ NO or NC contact Nylon Housing M12 Connector Optional M12 Con...

Страница 7: ...er Supply NC Contact P1 Set Point Potentiometer Adjust B1 8 LED s Red LED Yellow LED Green LED L1 Central LED Green With flow Red No flow Bargraph 1 2 3 Ground 4 NO Contact 5 Common 6 Power Supply Pow...

Страница 8: ...larm LED Status RED GREEN Set Point OFF ON FSW SPDT Status NO 4 NC 6 C 5 C 5 Application FSL Condition Normal Alarm LED Status GREEN RED Set Point ON OFF FSW SPDT Status NO 4 NO 4 NO 4 NC 6 NC 6 NC 6...

Страница 9: ...e risk of air accumulating in the pipe Keep in mind that the ideal mounting location is where the pipe is always full This will ensure that the switch is always immersed in the flow Fig 2 correct fig...

Страница 10: ...on the side as long as the tip of the probe sits within the middle of the pipe See Fig 2 When the pipe is in a vertical position the FSW should be installed only when the water flow is upward See Fig...

Страница 11: ...ter clockwise until the central LED changes to a blinking green state 4 Continue to turn the potentiometer clockwise until all green LEDs in the bar graph are on 5 In order to be sure that the adjustm...

Страница 12: ...for 2 minutes 2 Turn the potentiometer clockwise until the central LED is blinking green 3 Turn the potentiometer counter clockwise until the central LED turns red 4 Continue to turn the potentiometer...

Страница 13: ...rosion or deposit build up If deposits are found clean the sensor to ensure optimum performance The FSW should not be dropped or suffer any impact or fall that could damage the electronics or the ther...

Страница 14: ...ousing 2 SPDT Large Aluminum Housing Cable gland w 2000mm cable M12 connector or NPT Red led flow is below setpoint Yellow led flow is at setpoint Green led flow is above setpoint 1450 PSI 100 Bar 450...

Страница 15: ...low is above setpoint Max Pressure 1450 PSI 100 Bar 4500 PSI 300 Bar upon request Accuracy Response Time Enclosure Material Glass filled nylon option Aluminum Wetted Material 316 Stainless Steel Class...

Страница 16: ...stallation insertion length Verify the calibration Check power supply Cause Solution Radio frequency interference Sensor is potentialy defective Flow switch turns on or off suddenly Relay remains clos...

Страница 17: ......

Страница 18: ...M 3993 1014...

Отзывы: