1-2
CAUTION & SAFETY INFORMATION
If the equipment is used in a manner not specified in this manual, the
protection provided by the equipment may be impaired.
The installation category is one (I).
The output terminals of this product are for use with equipments (Digital
meters, chart recorders, etc.) which have no accessible live parts. Such
equipment should comply with all the applicable safety requirements.
Avoid contact with Hazardous live parts.
Do not operate the equipment in flammable or explosive environments.
The equipment comes with a 4.6 m (15') shielded multi-conductor power/
output cable. This cable is an ten conductor 26 AWG stranded wire cable with
rating of 300 Vdc, 105ºC (221ºF) PVC insulation. It also comes with a 2m (6')
shielded USB + Four conductor cable.
Power must be disconnected before making any electrical connections.
The supply voltage to the equipment should not exceed 24 Vdc.
The recommended power supply should be VDE or UL approved. Rating of 15
to 24 Vdc @ 150 mA minimum power with overload protection, current limited
to 500 mA.
There is no user replaceable fuse in this product.
The FMA1000 series passed all the EMC standards (EN61326 Emissions &
EN61000 immunity) under normal mode of operation using the power/output
cable provided and an external DC power supply.