5.2 Error Code and Counter-Measures
The FDT-25W ultrasonic flow meter will show Error Code in the lower right corner with a single letter
like I, R etc. on menu windows M00, M01, M02, M03, M90 and M08. When any abnormal Error Code
shows, counter-measures should be taken.
Error
code
Correspondent Message
displayed on M08
Causes
Counter-measures
R
System Normal
No error
I
Detect No Signal
(1)No Signals detected
(2)Transducers installed
improperly
(3)Too much fouling
(4)Pipe liners are too thick.
(5)Transducer cords are not
properly connected
(1)Relocate
measuring
location
(2)Clean the spot
(3)Check the cords
J
Hardware Error
Hardware problem
Contact the factory
H
PoorSig Detected
(1)Poor signal detected
(2)Transducers installed
improperly
(3)Too much fouling
(4)The pipe liners are too thick.
(5)Problem with transducers
cords
(1)Relocate measuring place
(2)Clean the spot
(3)Check the cords
(4)Check the coupler
Q
Frequ Output Over
The actual frequency for the
Frequency Output is out of the
range set by the user
Check the value entered at
M66, M67,M68 and M69, and
try to enter a larger value on
M69
F
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
(1) Temporary problems with
RAM, RTC
(2) Permanent problems with
hardware
(1) power on again
(2) contact factory
G
Adjusting Gain
Instrument is in the progress of
adjusting the gain for the
signal,
and
the
number
indicates the progressive steps
K
Empty pipe
No liquid inside pipe
Setup error on M29
Relocate where the pipe is
full of liquid
Enter 0 on M29