Section 5 - Mounting
The EWS Series of sensors/transmitters are designed for wall mounting in locations that
are free from dirt, grease, food particles and condensing moisture such as manufacturing
clean rooms, computer rooms and laboratory type environments. Wall anchors and
mounting screws are included for mounting. A conduit box mounting bracket/wall plate
adapter kit is available that will allow the transmitter to be mounted to a standard
electrical conduit box.
Section 6 - Transmitter Wiring Examples
For current output (4-20 mA)
For voltage output (1-5 Vdc)
3
COVER PLATE
GROMET
ADAPTER PLATE
CONDUIT BOX (NOT INCLUDED) IS SHOWN FOR
REFERENCE ONLY.
MULTIMETER
POWER SUPPLY
P
+
+
–
–
1 2
3
1
TB1
16.80 mA
MULTIMETER
POWER SUPPLY
+
+
+
–
–
1 2
3
T
4.20 Vdc
1
Order OMEGA
®
Model No.:
EWS-MB
.