DVG-64

Electronic Vacuum Gauge

M3805/0117

1

INSTRUCTION

SHEET

Shop online at omega.com

SM

   e-mail: [email protected] 

For latest product manuals: www.omegamanual.info

DVG-64 SPECIFICATIONS

Display 

Updated every second. 1 micron resolution below 200 micron of vacuum

Sensor T ype 

Thermistor 

Connector  T ype 

Standard 1/4 inch male flare fitting 

V acuum  Range 

0 - 12,000 Microns (0 - 1,600 Pascals) with vacuum increasing/decreasing indicator when above  

12,000 Microns  

Scale 

Microns, PSI, InHg, milliBars, Pascals, Torr, milliTorr 

Operating  Temp. Range  2 to 52°C (35 to 125°F) 
Overpressure 

300 PSI max (20 Bar) 

Accuracy 

±10% (0 to 1000 microns)

P ower  Source 

9 V olt Alkaline battery (not included)

Continuous  Usage 

Over 35 hours 

Auto Shut off 

After 10 minutes when vacuum reading is above 12,000 Microns (12 T orr) 

W eight 
Dimensions 

FEATURES 

• Reads vacuum in 7 international units 
• Uses  the battery power than other models 
• Cleaning port provides quick cleaning and drying of sensor 
• Uses standard 9V Alkaline battery 
• Convenient built-in hanger 

OPERATING CONTROLS 

• Turning the vacuum gauge ON: Press and hold the ON button  
 for about 3 seconds, until "POWER UP" appears on the display. 
• Turning the vacuum gauge OFF: Press the OFF button 
  To prolong battery life, when vacuum reading is above 12,000 
  Microns for about 10 minutes, the unit will automatically turn OFF. 
• Changing the scale: Press the Scale button to change the display 
  to the next scale. The scale order is: Microns, PSI, Inches of  
  mercury (InHg), milliBars, Pascals, Torr and milliTorr. DVG64 will  
 keep the scale settings even if the power is turned OFF. 

  

UNDERSTANDING THE DISPLAY 

• When the vacuum reading is above 12,000 Microns (1,600  
 Pascals),  the first line of the display shows "Atm." The second  
  line displays a bar graph to indicate the direction in which the  
 vacuum is moving.  When the bar graph is moving from left to  
 right, the pressure is increasing. When the bar graph is moving  
  from right to left, the pressure is decreasing. The speed of the 
  bar graph indicates how fast the pressure is increasing or  
  decreasing. The bar graph indicator may be inaccurate for a few  
  seconds after the evacuation of the system has begun. 

• The bar graph disappears if the vacuum does not change for  
  about 10 seconds. When the vacuum reading is below 12,000 
  Microns (1,600 Pascals), the vacuum in the selected units is  
  displayed. 

CONNECTING TO THE VACUUM SYSTEM 

The VG64 should only be connected to the vacuum system at   
the vacuum port. The "Auxiliary Port" is primarily for cleaning   
and should normally be closed with the supplied cap. 

CLEANING THE VACUUM SENSOR 

It is recommended that the sensor be cleaned periodically to  
maintain unit accuracy . Oil and other contaminants reduce the  
accuracy of the unit. Follow the instructions below for cleaning. 

• Close the Vacuum port with the supplied cap.  
  Open the Auxiliary port. 
• Use an eyedropper to pour about 2 teaspoons of ordinary  
  rubbing alcohol into the Auxiliary port. 
• Close the Auxiliary port with the supplied cap. Both the  
 Vacuum and Auxiliary ports should now be closed. 
• Shake the DVG64 unit for about 10 seconds. A slight  
  movement of the vacuum sensor in the case is normal and  
  does not affect the internal connection in any way. 
• Open both the Vacuum and the Auxiliary ports. 
  Empty the alcohol, and air dry the sensor. 
• Close both the Vacuum and the Auxiliary ports with the  
  supplied caps when the DVG64 is not used. This prevents  
  contamination of the sensor. 

V acuum (microns) 

Micron  (mT orr)  Step 

P ascal Step 

milliBar Step

InHg Step 

PSI Step 

15000 - 8000 

1000 

100 

0.05 

0.02 

8000 - 5000 

500 

5 0 

0.5 

0.02 

0.01 

5000 - 2000 

250 

2 5 

0.25 

0.01 

0.005 

2000 - 1000 

5 0 

0.05 

0.002 

0.001 

1000 - 500 

1 0 

0.01 

0.0005 

0.0002 

500 - 200 

0.5 

0.0005 

0.0002 

0.0001 

200 - 0 

0.1 

0.0001 

0.00005 

0.00002 

DVG64 STEP TABLE 

   

0.2 kg (6.7 oz)

140 H x 76 W x 32 mm D (5.5 x 3 x 1.25")