
Installation, Mounting & Wiring
4
4-3
• Immediately discontinue use of the battery if the battery emits an unusual
smell, feels hot, changes color or shape, leaks or appears abnormal in any
other way.
• Do not place the battery in microwave ovens or high-pressure containers.
Standard 1.5V C cell Alkaline batteries can be used as
an alternative to the standard Lithium batteries. You must
first remove all batteries from the gauge. Then configure
jumper S2 with only position M installed as indicated
in Figure 4-3. Always remove all 3 of the S2 jumpers
before changing configurations. Failure to properly
configure the jumpers can damage the gauge. Battery
life is subject to usage conditions. Battery life estimates
are provided for Lithium batteries only.
Figure 4-3. Battery Type Configuration
4.5 Wiring (Power, Analog Output, Alarm)
Power Supply Wiring Example
Figure 4-4. Wiring - Power Supply
TB2
PIN 1
5
4
24Vdc POWER SUPPLY
+
–
TB2
TB1
BATTERY
BATTERY
S2-D&E
M
E
D
ALKALINE SETTING*
LITHIUM SETTING*
S2-M
E
D
M
*IMPORTANT SEE NOTE!
NOTE:
Содержание DPG409 SERIES
Страница 6: ...iv NOTES ...