2
About The Meter
6
2.4 FRONT OF THE METER(Continued)
METER BUTTONS
SETPTS
Button
In the Run Mode, this button will sequentially recall the previous setpoint settings. As
necessary, use the
/MAX
and
/TARE
buttons to alter these settings, then press the
SETPTS
button to store new values.
Unless you press the
SETPTS
,
/TARE
, or
/MAX
button within 20 seconds, the meter
will scroll to setpoint 2 and then to the Run Mode.
If the dual relay option is not installed or if the
L.3=1
on the
LK.CF
menu,
pressing the
SETPTS
button will display the meter's firmware version.
/MAX
Button
In the Run Mode, this button will recall the
PEAK
reading since the last press of the
RESET
button.
In the Configuration Mode, press this button to change the value of the flashing digit
shown on the display and/or toggle between menu choices, such as
R.1=T
or
R.1=N
on
RD.CF
menu. When configuring your setpoint values, press the
/MAX
button to
advance the flashing digit's value from 0 to 9 by 1.
/TARE
Button
In the Run Mode press the
/TARE
button to tare your reading (zeroing) if you
configure the Reading Configuration bit
R.1=T
of the
RD.CF
menu. If you configure
R.1=N
, the
/TARE
button has no function.
In the Configuration Mode, press the this button to scroll to the next digit.
Note
☞
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание DP25-E
Страница 11: ...viii Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 70: ...59 NOTES Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...