Page 13
General
Accuracy
±(0.025% of R 0.05 Ohms)
Temperature Drift
± 0.01% of span outside of 23°C ±10 °C (73°C ±18 °F)
Operating Temperature Range
-25 to 60 °C (-10 to 140 °F)
Relative Humidity Range
10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
Size
5.63 x 3.00 x 1.60 inches, 143 x 76 x 41 mm (L x W x H)
Weight
12.1 ounces, 0.34 kg (including boot & batteries)
Batteries
Four “AA” Alkaline 1.5V (LR6)
Battery Life
50 Hours
Optional NiMh Rechargeable battery kit
120 VAC for North America Only; charger, four NiMh
batteries, AC & DC cords [Part # 020-0103]
Low Battery
Low battery indication with nominal 1 hour of operation
left
Protection against misconnection
Over-voltage protection to 60 V dc (rated for 30 seconds)
Display
High contrast graphic liquid crystal display. LED backlighting
for use in low lit areas.
Read
Excitation Current
0.9 mA to 401 Ohms, 0.4 mA to 2010 Ohms (nominal)
Normal Mode Rejection
50/60 Hz, 50 dB
Common Mode Rejection
50/60 Hz, 120 dB
(Unless otherwise indicated all specifications are rated from a nominal 23 °C,
70 % RH for 1 year from calibration)
Source
3 Wire & 4 Wire Accuracy
From 1 to 10.2 mA External Excitation Current
Below 1 mA of External Excitation Current
2 Wire Accuracy
±(0.025% of R 0.05 Ohms)
±(0.025% of R + 0.05 Ohms)
Add 0.1 Ohms to 3 Wire & 4 Wire Accuracy
Resistance Ranges
0.00 to 401.00, 0.0 to 2010.0 Ohms
Allowable Excitation Current Range
400 Ohm Range:10.2 mA max; steady or pulsed/intermittent
2000 Ohms Range: 1 mA max; steady or pulsed/intermittent
Pulsed Excitation Current Compatibility
DC to 0.01 second pulse width
0.025 mV
mA Excitation Current
CL514-PLUS Specifications