9
10. Specifications
(1 year at 23°C ±5°C; % of reading unless otherwise noted)
Output Voltage
Range
-10 to +75.000 mV
Resolution 1
μ
V
Accuracy
±0.007 % of reading, ±5
μ
V
Output Impedance
>1 Ohm
Input Voltage
Range
-10 to +75.000 mV
Resolution 1
mV
Accuracy
±0.007 % of reading, ±5
μ
V
Input Impedance
> 1 MegOhm
Thermocouple Source/Measure
Types
J, K, T, E, R, S, B, L, U, C, mV
Resolution
Source
±0.1°C or °F
Measure
±0.01°C or °F
Accuracy
J
±0.5 °C; -210 °C to -100 °C
±0.25 °C; -100 °C to +1,200 °C
K
±0.6 °C; -200 °C to -100 °C
±0.3 °C; -100 °C to +1,000 °C
±0.5 °C; +1,000 °C to +1,372 °C
T
±0.7 °C; -200 °C to -150 °C
±0.3 °C; -150 °C to +400 °C
E
±0.5 °C; -200 °C to -100 °C
±0.2 °C; -100 °C to +1,000 °C
R
±1.8 °C; 0 °C to 250 °C
±1.0 °C; 250 °C to +1,767 °C
S
±1.8 °C; 0 °C to 250 °C
±1.0 °C; 250 °C to +1,767 °C
B
±1.7 °C; 600 °C to 1,000 °C
±1.2 °C; 1,000 °C to 1,820 °C
L
±0.5 °C; -200 °C to -100 °C
±0.4 °C; -100 °C to +900 °C
U
±0.7 °C; -200 °C to 0 °C
±0.3 °C; 0 °C to +600 °C
C
±0.4 °C; 0 to °C 1,000 °C
±0.7 °C; 1,000 °C to +1,800 °C
±1.2 °C; +1,800 °C to +2,316 °C
CJC Temperature Offset
±0.05 °C/°C outside of 23±5 °C
Warm-up Time
1 minute to specification
Environmental
Operating
-10 °C to +55 °C
Storage
-20 °C to +70 °C
Power Requirements
Battery
9 Volt alkaline; 006P/ IEC 6F22/
NEDA1604
Optional NiCad
Optional AC adapter/charger
Mechanical Dimensions
5.7" x 3.15" x 1.43"
(14.47 cm x 8.00 cm x 3.63 cm)
Weight
12 ounces (340 grams)
Notes: 1.Temperature standard ITS-90.