Mounting

Mount the unit on a bench, table top or shelf in a 

horizontal position and operate at least ten inches 

from any air obstructions to the fan, front, rear 

panel, bottom and top of the unit, in an ambient 

environment between the specified 0 to 50°C  

(32 to 122°F).

Your unit was shipped to you with a wood plug 

installed in the heater cavity to prevent damage dur-

ing shipping. This plug must be removed before using 

your calibrator.

Ambient Temperature

The target plate of the BB-4A can achieve any 

temperature within the specified temperature range 

when being operated in normal ambient temperature 

environments. The maximum specified heater cavity 

temperature of 982°C (1800°F) can be achieved over 

the entire specified ambient temperature range.

Front Panel

Process Temperature:

This field displays the current temperature of the 

target plate.

Setpoint Temperature:

This field displays the desired target plate 

temperature. Once the target plate reaches this 

desired temperature, both displays will read the same 

value.

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

START HERE

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

2

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

3

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

4

Using This Quick Start Manual

Use this Quick Start Manual with your BB-4A 

Blackbody Calibrator for easy installation and 

basic operation. For detailed information, refer to 

the User’s Guide (Manual # M3264).

PRECAUTIONS:

•  Follow all safety precautions and operating 

instructions outlined in this quick start and 

accompanying User’s Guide.

•  Never leave your calibrator unattended when in 

use.

•  Keep out of reach of all children.
•  Remove the cavity wood shipping plug before use.
• Never touch the heater cavity when hot.
•  Never place any object within 3 inches of the 

cavity opening when hot.

•  Do not operate in flammable or explosive 

environments.

•  Never operate with a power cord other than the 

one provided with your unit.

•  Remove and or disconnect main power cord 

before attempting any maintenance or fuse 

replacement.

•  Do not connect and or operate this unit to a 

non-grounded, non-polarized outlet or power 

source.

•  Do not connect the serial port or reference 

probe port to equipment with exposed, 

hazardous, live voltages.

• Protect from moisture and rain.
•  The operator of this instrument is advised that if 

the equipment is used in a manner not specified 

in this manual, the protection provided by the 

equipment may be impaired.

General Information

The Model BB4A is a portable, rugged, bench-

top, blackbody calibrator source with a built-in 

precision PID digital controller. The calibrator is 

used to test and calibrate infrared pyrometers. 

The heater cavity, with 21.66 mm (0.88") aperture, 

has an emissivity of 0.99 and can be set to any 

temperature between 100 to 982°C (212 to 1800°F).

 

Parameter/Access Key:

  

Press to scroll through menu parameters.

  

Raise Key:

  

Press to increase the selected parameter or scroll 

upward in the list of possible settings.

  

Lower Key:

  

Press to decrease the selected parameter or scroll 

downward in the list of possible settings.

    

Mode Key:

 

Press to save settings and exit a menu level.

Back Panel

AC Power Mains Input and Fuse Compartment: 

 

The customer connects the power cord to the AC 

Power Input. As a safety precaution, the power cord 

cannot be connected if the fuse compartment is open. 

Refer to the User’s Guide for information on fuse 

replacement.

Reference Probe Connection: 

 

The reference probe enables the user to monitor the 

target plate temperature with an external instrument. 

A “K” type thermocouple reference output is 

provided.

RS232 Communications Port: 

 

The Female DB-9 port allows the customer to make a 

3-wire RS232 interface with the BB-4A.

Thermal Reset Switch

 

In the event that the target plate temperature 

and/or ambient temperature exceed the 

acceptable limit, this reset switch will pop open. 

The controller will appear to be demanding 

a temperature increase but the target plate 

temperature will not rise. Let the unit cool down, 

completely. Press the switch located on the right 

side of the unit, behind the carry handle until a 

click is heard. If this has no effect, refer to the 

troubleshooting section in the User’s Guide.  

Changing the Temperature Setpoint

 

The BB-4A’s upper display indicates the 

blackbody target plate temperature known as 

(PV) Process Variable, while the lower display 

indicates the programmed setpoint known as 

(SV) Setpoint Variable. Making changes to the 

setpoint, units of measure (°F/°C) and 

communication settings (BAUD, etc.) are made 

via the 

 and 

 keys.

 Holding a key in, continuously, will cause the 

setpoint temperature to advance more quickly 

to a desired value. Three scanning speeds are 

provided: slow, medium and fast. The minimum 

and maximum setpoints are locked.

Menu Heat Up/Cool Down Cycle Time Tables

Approximate cycle times for heat up and cool 

down are given in the table above. To find a 

given transition time from an initial temperature 

to a second target temperature follow this 

procedure: Look for an initial temperature in the 

left column. Next, look for the target temperature 

along the top row. The intersection of the row 

and column provides the approximate transition 

time. 

BB-4A

212F/100C

Temperature

Range

1800F/982C

TM

SETPOINT TEMPERATURE
PROCESS TEMPERATURE

HEATER CAVITY

MODE KEY

TEMPERATURE RANGE
REFERENCE BAR

LOWER SETPOINT KEY
RAISE SETPOINT KEY

PARAMETER/

ACCESS KEY

ON/STAND-BY

POWER SWITCH

AC MAINS POWER INPUT
AND FUSE COMPARTMENT

TC REFERENCE
PROBE CONNECTION

FAN

FUSE

RS-232 COMMUNICATIONS PORT

NOTE:

  

 

121ºC 

538ºC 

982ºC

121ºC 

 

25 min.  40 min. 

250ºF

538ºC 

60 min.  30 min.  1000ºF

982ºC  140 min.   30 min.  1800ºF 
   250ºF 

1000ºF  1800ºF 

TO:

TO:

FROM:

FROM: