background image

CABLE TYPE F2

Red

=

+ Power

White

=

Common

Green

=

Output

CABLE TYPE C1

Red   

=  + Power

White =  Common

Green =  Output

HIRSCHMANN TYPE 

PIN-1

=

+ Power

PIN-2

=

Common

PIN-4

=

Output

BENDIX 4-PIN, 6-PIN

Pin A

=

+ Power

Pin B

=

Output

Pin D

=

Common

Voltage Output Units 1-5, 1-6 Vdc

Current Output Units 4-20mA

Ratiometric (mV/V)

PX791 & PX792 SANITARY TRANSDUCERS – 

ELECTRICAL CONNECTIONS

CABLE TYPE F2

Red

=

+ Power

White

=

– Power

Green

=

+ Output

Black

=

– Output

CABLE TYPE C1

Red   

+ Power

White =

– 

Power

Green =

+ Output

Black

=

– Output

HIRSCHMANN TYPE 

Pin A

=

+ Power

Pin B

=

+ Output

Pin C

=

– Output

Pin D

=

– Power

BENDIX 4-PIN, 6-PIN

Pin A

=

+ Power

Pin B

=

– Power

Pin C

=

+ Output

Pin D

=

– Output

Pin E

=

Shunt Cal.

Pin F

=

Shunt Cal.

CABLE TYPE F2

Red

=

+ Power

Black

=

Power

CABLE TYPE C1

Red

=

+ Power

Black

=

Power

HIRSCHMANN TYPE 

Pin-1

=

+ Power

Pin-2

=

Power

BENDIX 4-PIN, 6-PIN

Pin A

=

+ Power

Pin B

=

Power

WA R N I N G :

R E A D   B E F O R E

I N S TA L L AT I O N

The diaphragm is very sensitive and fragile!
Do not let anything touch the diaphragm but
the fluid to be measured. Fluid hammer and
surges can destroy any pressure transducer
and must always be avoided. A pressure
snubber should be installed to eliminate the
damaging hammer effects. Fluid hammer
occurs when a liquid flow is suddenly
stopped, as with quick-closing solenoid
valves. Surges occur when flow is suddenly
begun, as when a pump is turned on at full
power or a valve is quickly opened.

Liquid surges are particularly damaging to

pressure transducers if the pipe is originally
empty. To avoid damaging surges, fluid lines
should remain full (if possible), pumps
should be brought up to power slowly, and
valves opened slowly. To avoid damage
from both fluid hammer and surges, a surge
chamber should be installed.

Symptoms of fluid hammer and surge's

damaging effects:

a) Pressure transducer exhibits an output

at zero pressure (large zero offset). If
zero offset is less than 10% FS, user
can usually re-zero transducer, install 
proper snubber and continue 
monitoring pressures.

b) Pressure transducer output remains

constant regardless of pressure.

c) In severe cases, there will be no

output.

Содержание PX791 Series

Страница 1: ...www omega com e mail info omega com omega comTM User sGuide PX791 PX792 Series Sanitary Pressure Transducer...

Страница 2: ...TEL 33 0 130 621400 FAX 33 0 130 699120 Toll Free in France 0800 4 06342 e mail france omega com Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056 3017 FAX 49 0 7056 8540 T...

Страница 3: ...serve the following Ground the body of the transducer BEFORE making any electrical connections When disconnecting remove the ground LAST NOTE The braided shield and drain wire in the cable if supplied...

Страница 4: ...in the labeling can be reasonably expected to result in a significant injury to the user 2 A critical component is any component of a life support device or system whose failure to per form can be re...

Страница 5: ...iaphragm but the fluid to be measured Fluid hammer and surges can destroy any pressure transducer and must always be avoided A pressure snubber should be installed to eliminate the damaging hammer eff...

Страница 6: ...UCER POWER OUTPUT POWER SUPPLY Common Power Output METER TRANSDUCER 3 Wire Voltage 4 20mA POWER SUPPLY Power Power Output Output METER TRANSDUCER RED WHITE GREEN BLACK 4 Wire Ratiometric mV V WIRING D...

Страница 7: ...all not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA is...

Страница 8: ...ty Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Отзывы: