5.  Accessories

6.  Specifications

PH-2720-PA pH/ORP Preamplifier

Housing material:

CPVC

Compatible sensors:

PHE-3271

Flat Surface pH Electrode

PHE-2716

Bulb pH Electrode

ORE-2715

Flat Surface ORP Electrode

ORE-2717

Bulb ORP Electrode

Input Impedance:

>10

11

 

Operating temp.:

0 to 80 

°

C

Gain:

X1 (unity)

Input power:

±

4.5 to 

±

8 VDC, dual supply

Current consumption: <1 mA, dual supply
Quality standard:

CE

Dimensions:

PHE Series pH Electrodes - ORE Series ORP Electrodes

General Specifications

Wetted parts:
Sensor body:

CPVC

O-rings:

Viton®

Electrode junction:

Porous UHMW polyethylene

Quality standard:

CE

Maximum pressure/temperature ratings:
• 7 bar (100 psi) max. @ 

65 

°

C (149 

°

F)

• 4 bar (58 psi) max. @ 

85 

°

C (185 

°

F)

50

3

PSI

BAR

50 75

25

122 167

77

25

2

75

5

100

7

100

212

°

F

°

C

SAFETY INSTRUCTIONS FOR IN-LINE ELEC-
TRODE INSTALLATION
1. Do not remove from pressurized lines.
2. Do not exceed maximum temperature/pressure

specifications.

3. Do not install/service without following

installation instructions (see sensor manual).

4. Wear safety goggles and faceshield during

installation/service.

5. Do not alter product construction.
6. Failure to follow safety instructions could result

in severe personal injury!

CAUTION!

When replacing O-rings, apply O-ring lubricant to sensor O-rings
prior to preamplifier/electrode assembly.  Unlubricated O-rings
may score the preamplifier sealing surface.

7.  Preamplifier Troubleshooting

Procedure:
A. Install sensor adapter into preamplifier
B. With preamplifier and instrument connected, simulate pH 4, 7, and

10 and record displayed mV readings (right) and approximate
preamplifier response time between simulations.

C. Refer to table on next page for preamplifier troubleshooting tips.

Viton®
EPR 
Kalrez

PP

Material

FPP-1220-0021
FPP-1224-0021
FPP-1228-0021

PH-31542-TC

Part no.

Sensor 
cap

Description

Sensor O-ring,
2 required

3.6 in./

91 mm

1.8 in./

46 mm

3/4 in. NPT

female threads

Std.15 ft/

4.6 m cable

Simulator

Actual displayed

Input

mV

4.00, +177 mV

________________________

7.00, 0 mV

________________________

10.00, -177 mV

________________________

Preamplfier response time:  

________________________

Содержание PH-2720-PA

Страница 1: ...www omega com e mail info omega com User sGuide Shop online at PH 2720 PA Preamplifier...

Страница 2: ...ance USA and Canada Sales Service 1 800 826 6342 1 800 TC OMEGA Customer Service 1 800 622 2378 1 800 622 BEST Engineering Service 1 800 872 9436 1 800 USA WHEN TELEX 996404 EASYLINK 62968934 CABLE OM...

Страница 3: ...n Blue Green White Red Black mV Input Iso Gnd T T V V OMEGA PHCN 5700 OMEGA ORCN 90 ORP CONTROLLER ENTER MOD CAL RELAY OUTPUT ORP mV RLY 1 RLY 2 RLY 3 RLY 4 Red Black Green White Brown Blue 10 V 11 V...

Страница 4: ...rocess pipe 60 Customer supplied 3 4 in NPT threaded pipe A J box B Flex conduit C Quick release pipe clamp D Tank flange E A B C C E B A D Closed tanks Open tanks In Line Caution If liquid level is n...

Страница 5: ...stallation instructions see sensor manual 4 Wear safety goggles and faceshield during installation service 5 Do not alter product construction 6 Failure to follow safety instructions could result in s...

Страница 6: ...sensor tip in pH 4 0 buffer during system maintenance intervals In line applications should be plumbed with a depression trap which ensures liquid is maintained around the sensor tip If sensor dehydr...

Страница 7: ...shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA...

Страница 8: ...ers Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pum...

Отзывы: