background image

Ref.No.

022015 

 

 

 

CAUTION 

1.

 

WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE TWO LASER BEAMS–PERMANENT EYE DAMAGE MAY RESULT. 

2.

 

USE EXTREME CAUTION WHEN OPERATING THE LASER. 

3.

 

NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.   

4.

 

KEEP OUT OF REACH OF ALL CHILDREN. 

STORAGE & CLEANING       

It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care 

should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the 

thermometer. Do not submerge any part of the thermometer.   

LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic messages as follows: 

 

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and LAL. 

 

 

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed 
when the ambient temperature exceeds 0

o

C (32°F) or +50

o

C (122°F). The thermometer should be allowed plenty of 

time (minimum 30 minutes) to stabilize to the working/room temperature. 

 

Error 5~9

, for all other error messages it is necessary to reset the thermometer. To reset it, wait for auto power off, 

remove  the  battery  and  wait  for  a  minimum  of  one  minute,  reinsert  the  battery  and  turn  on.  If  the  error  message 

remains please contact the Service Department for further assistance. 

」「

 

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.

 

BATTERIES

 

The thermometer incorporates visual low battery indication as follows: 

 

 

 

‘Battery OK’:   

measurements are possible 

‘Battery Low’: battery needs to be   

replaced, measurements are still possible 

‘Battery Exhausted’:   

measurements are not possible 

  When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA x 2 batteries.   

Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction. 

  Dispose of used battery promptly and keep away from children. 

SPECIFICATION 

Item 

Non-contact Infrared Scan function 

Thermocouple Probe Scan function 

(K type; a bead wire included.) 

Measurement Range 

100 to +1800

 o

C (212 to +3272

 o

F)

 

K Type: -64 to +1400°C (-83.2 to +2552°F) 
Bead : -50 to +250°C(-58 to +482°F) 

Operating Range 

0 to +50°C (32 to +122°F) 

Accuracy (Tamb=23±3°C) 

±2% of reading 

±1% of reading or 1°C (1.8°F) whichever is 

greater (Test under Tamb=23±6

°

C) 

Emissivity Range 

0.95 default – adjustable 0.1 to 1 step .01 

Resolution   

0.1°C/0.1°F at -83.2 to 999.9(°C/°F), otherwise 1°C/1°F 

Response Time (90%) 

1 sec. 

Distance:Spot 

75:1 

Memory 

Detailed 24 Memories with Temperature & Emissivity 

USB Function 

Please download the program from CD ( enclosed with the package) 

Battery Life 

Typ.100, min70 hours continuous use (Alkaline, without Laser and Back Light.) 

Dimensions 

204.14*179.19*58.40 mm (8.04×7.05×2.30 inch) 

Weight 

355 grams(12.52 oz) including batteries (AAA*2pcs) 

 

EMC/RFI: 

Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per meter, 

but the performance of the instrument will not be permanently affected.

 