Ref.No.
:
022015
CAUTION
1.
WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE TWO LASER BEAMS–PERMANENT EYE DAMAGE MAY RESULT.
2.
USE EXTREME CAUTION WHEN OPERATING THE LASER.
3.
NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4.
KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care
should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the
thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and LAL.
「
」
「
」
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ is displayed
when the ambient temperature exceeds 0
o
C (32°F) or +50
o
C (122°F). The thermometer should be allowed plenty of
time (minimum 30 minutes) to stabilize to the working/room temperature.
「
」
Error 5~9
, for all other error messages it is necessary to reset the thermometer. To reset it, wait for auto power off,
remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message
remains please contact the Service Department for further assistance.
「
」「
」
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’:
measurements are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’:
measurements are not possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA x 2 batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Item
Non-contact Infrared Scan function
Thermocouple Probe Scan function
(K type; a bead wire included.)
Measurement Range
100 to +1800
o
C (212 to +3272
o
F)
K Type: -64 to +1400°C (-83.2 to +2552°F)
Bead : -50 to +250°C(-58 to +482°F)
Operating Range
0 to +50°C (32 to +122°F)
Accuracy (Tamb=23±3°C)
±2% of reading
±1% of reading or 1°C (1.8°F) whichever is
greater (Test under Tamb=23±6
°
C)
Emissivity Range
0.95 default – adjustable 0.1 to 1 step .01
Resolution
0.1°C/0.1°F at -83.2 to 999.9(°C/°F), otherwise 1°C/1°F
Response Time (90%)
1 sec.
Distance:Spot
75:1
Memory
Detailed 24 Memories with Temperature & Emissivity
USB Function
Please download the program from CD ( enclosed with the package)
Battery Life
Typ.100, min70 hours continuous use (Alkaline, without Laser and Back Light.)
Dimensions
204.14*179.19*58.40 mm (8.04×7.05×2.30 inch)
Weight
355 grams(12.52 oz) including batteries (AAA*2pcs)
EMC/RFI:
Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per meter,
but the performance of the instrument will not be permanently affected.