background image

1)

 

Introduction 

 

The OS1300-Series handheld infrared thermometer is a non-contact temperature sensing thermometer. It has a wide 

temperature range of -50 to 1300°C (-58 to 2372°F). This thermometer features a backlit display which is automatically 

illuminated when the trigger is pulled and allows working in poor or zero-lighting conditions. The single laser is also 

automatically active when the trigger is pulled. The laser sight can be turned on and off with the press of a button. The laser 

is located above the IR sensor to locate the center of the area being measured. The field of view is 16:1. The thermometer 

constantly updates the temperature measurement as long as the trigger is held in and will “hold” on the last temperature 

seen when the trigger is released. The OS1300-Series has adjustable emissivity, min, max, and averaging functions, as well 

as high and low alarm capability. This safely measures surface temperatures of hot, hazardous, or hard-to-reach objects 

without contact. 

 

1.1 How it works 

This handheld infrared thermometer measures the surface temperature of an object. The unit’s optics sense emitted, 

reflected, and transmitted energy which is collected and focused onto a detector. The unit’s electronics converts the energy 

value into a temperature in °C or °F which is displayed on the digital display. A clean line of sight free of dust or mist is 

needed between the sensor and the object. For increased ease of use and accuracy the laser pointer makes aiming even 

more precise. 

 

 

Measurement: 

When taking measurement, point the thermometer toward the object to be measured and hold the trigger.  The 

temperature will be displayed and the word “SCAN” will flash on the left-hand side of the LCD screen to indicate that the 

temperature is continually being updated on the display. When the trigger is released, the last reading seen will be held on 

the display and the word “HOLD” will be visible on the left-hand side of the LCD screen. The object under test should be 

larger than the spot size calculated by the field of view diagram below. 

 

 

Field of View: 

Field of view, or distance to spot ratio, is the size of an area that can be measured from a specified distance. As the distance 

from the object increases, the spot size of the measuring area becomes larger. Make sure the target is larger than the unit’s 

spot size. When accuracy is critical, ensure that the target is at least twice as large as the spot size. Please see field of view 

diagram for the OS1300-Series below. 

 

 

 

 

Emissivity: 

Most organic materials and painted or oxidized surfaces have an emissivity of 0.95 (pre-set in the unit but adjustable). 

Inaccurate readings will result from measuring shiny or polished metal surfaces. To compensate, cover the measured 

surface with masking tape of flat black paint. Measure the tape or painted surface when the tape or painted surface 

reaches the material temperature. Please see the Emissivity table below for common emissivity values and adjust 

accordingly. 

Содержание OS1300 Series

Страница 1: ...TM User sGuide Shop online at omega com e mail info omega com For latest product manuals www omegamanual info OS1300 Series Performance Handheld Infrared Thermometer...

Страница 2: ...ustomer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only U S A Headquarters Tel 203 359 1660 e mail info omega com Fax 203 359 7700 The information contained i...

Страница 3: ...ts Table of Contents 3 1 Introduction 4 1 1 How it works 4 1 2 Caution 5 2 Specifications 6 3 Operating Instructions 6 3 1 Battery Installation 6 3 2 Button and Key Layout 7 3 3 Locating a Hotspot 7 4...

Страница 4: ...e sensor and the object For increased ease of use and accuracy the laser pointer makes aiming even more precise Measurement When taking measurement point the thermometer toward the object to be measur...

Страница 5: ...ozen Food 0 90 Steel 0 80 Hot food 0 93 Textiles 0 94 Glass plate 0 85 Water 0 93 Ice 0 98 Wood 0 94 1 2 Caution Infrared thermometer should be protected for the following EMF electro magnetic fields...

Страница 6: ...ttery door back to its original position Figure 1 Figure 2 Figure 3 Temperature Range 50 to 1300 C 58 to 2372 F Accuracy 3 of reading 3 C 50 C to 0 C 58 to 32 F 2 of reading 2 C above 0 C Field of Vie...

Страница 7: ...own e LAL shows Low Alarm value use UP and DN buttons to adjust up or down f DIF shows realtime difference between current reading and original starting value 3 4 2 Changing Temperature Units 1 With t...

Страница 8: ...ometer outside the area of interest then scan across with up and down motions until you locate the hot spot Turning on the laser will make for more precise targeting 4 Maintenance 1 Lens cleaning blow...

Страница 9: ...r otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental orspecial damages CONDITIONS Equipment sol...

Страница 10: ...puters Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators...

Отзывы: