2)
Specifications
3)
Operating Instructions
3.1
Battery Installation:
Pull open the battery compartment
(1)
, connect 9V battery to the correct positive and negative terminals
(2)
, place
connected battery into battery compartment and close cover
(3)
.
Temperature Range
-50°C - 1550°C (-122°F - 2822°F)
Thermocouple Range
-50°C - 1370°C (-58°F - 2498°F)
Accuracy
±4.0°C (7.2°F): -50°C - 0°C (-58°F - 32°F)
±1.5% ± 2.0°C (3.6°F): 0°C - 500°C (32°F - 932°F)
±2.0% ± 2.0°C (3.6°F): >500°C (932°F)
Thermocouple Accuracy
±1.5% or 2.0°C (3.6°F), whichever is greater
Field of View
30:1
Repeatability
1% of reading or 1
°C
Response Time
<250 ms (95% response)
Spectral Response
8 – 14 µm
Emissivity
Adjustable emissivity 0.10 – 1.00
Operating Temperature Range
0 - 50°C (32 - 120°F)
Relative Humidity
10 -90% RH, Storage <80% RH
Storage Temperature
-20 - 60°C (-4 - 140°F)
Weight
286g
Dimensions
7.80 x 4.72 x 2.32 in (198 x 120 x 59 mm)
Record Memory
99 records
Diode Laser
Output < 1mW, 630-670nm, class 2 (II)
Power
9V Battery
Automatic Power Shutoff
Automatic shutoff after 25 seconds
Estimated Battery Life
16 Hours
1
2
3