LED FLASHING MODES

(Single green flash)

The channel is logging, no alarm

(Single r ed flash)

The channel is logging,  low alarm

(Double red flash)

The channel is logging, high alarm

The number of  flashes and colours indicates :

BATTERY REPLACEMENT

 

Note:

Leaving the OM-EL-USB-LITE plugged into the USB port for longer than necessary will cause some of the battery capacity to be lost.

WARNING:  Handle lithium batteries carefully.

Dispose of in accordance with local regulations.

Red/Green LED

Temperature sensor

We recommend that you replace the battery every 1 month, or prior to logging critical data. The OM-EL-USB-LITE does not lose 
its stored readings when the battery is discharged or when the battery is replaced; the data logging process will however be
stopped and cannot be re-started until the battery has been replaced and the logged data has been downloaded to PC.
For maximum logging life, use only Omega model no. OM-EL-BATT-LITE replacement battery. Before replacing the battery, 
remove the OM-EL-USB-LITE from the PC.

OM-EL-USB-LITE features a bi-colour LED to indicate the status of the data logger. To save power, the status indication only operates 
when the push button is pressed for a short while.