11. Calibration

Your transmitter has been factory calibrated to meet or exceed the
specifications outlined in this manual. To maintain original
specifications it is generally recommended that your transmitter be re-
calibrated on an annual basis depending on operating conditions.

12. Calibration Procedure for HX93AC (all styles)

Relative Humidity Adjustment

Recommended equipment:

• Humidity Calibration Kit, OMEGA

®

Model No.: 

HX92-CAL

• Handheld Digital Multi-meter, OMEGA

®

Model No.: 

HHM29

• Dc Power Supply, OMEGA

®

Model No.: 

PSU-93

1. Remove enclosure cover
2. Connect transmitter as shown in figure below.
3. Apply power to transmitter and allow to warm up for 15 min. 
4. Place unit in a 11% RH environment and allow to stabilize for 15 min.
5. Adjust potentiometer “P1” for a output reading of 5.76 mA
6. Place unit in a 75% RH environment and allow to stabilize for 15 min.
7. Adjust potentiometer “P2” for a output reading of 16.0 mA
8. Repeat steps 4, 5, 6, and 7 as necessary until proper readings are maintained.
9. Calibration complete.

Figure 6 - HX93AC Calibration Procedure

Relative Humidity Adjustment

16.00 mA

TB1

TB2

24V

PSU-93

HHM29

SENSOR HEAD

10A uAmA COM V

HX92-CA
CALIBRATION BOTTLE

P1 P3

P2

+

12

3

12

3

11

HX93A, HX93DA SERIES

RH/Temperature Transmitter

NOTE:

Do not adjust
P3. This is a
factory
setting.

Содержание HX93A SERIES

Страница 1: ...www omega com e mail info omega com User sGuide HX93A HX93DA SERIES RH Temperature Transmitter MADE IN Shop online at...

Страница 2: ...Guyancourt France Tel 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 0 7056 9398 0...

Страница 3: ...ulations 5 9 RH Measured Vs Output Reading 6 10 Temperature Output Calculations 6 11 Calibration 7 12 Calibration Procedure For HX93AC HX93DA Series Relative Humidity Adjustment 7 13 Calibration Proce...

Страница 4: ...mensions 3 Figure 4 Current Transmitter Wiring Example 4 20 mA 4 Figure 5 Voltage Transmitter Wiring Example 5 Figure 6 HX93AC Calibration Procedure Relative Humidity Adjustment 7 Figure 7 HX93AC Cali...

Страница 5: ...ing list and verify that you have received all your equipment If you have any questions about the shipment please call our Customer Service Department at 1 800 622 2378 or 203 359 1660 On the web you...

Страница 6: ...ignal measured is less susceptible to electrical noise interference and the loop can support more than one measuring instrument as long as the maximum loop resistance is not exceeded A typical applica...

Страница 7: ...unting depending upon model Plastic wall anchors and mounting screws are included for wall mounting and remote probe models A duct mounting kit is included with duct mount models Figure 2 Wall Mount M...

Страница 8: ...2 No Connection 2 No Connection 3 4 20 mA Output 3 4 20 mA Output 6 Current Transmitter Wiring Examples For current output models 4 20 mA Figure 4 Current Transmitter Wiring Example 4 20 mA All electr...

Страница 9: ...Example 8 Relative Humidity Output Calculations To calculate Relative Humidity by measuring the current or voltage output use the following formulas For current output RH Current measured in miliamps...

Страница 10: ...F EXAMPLE 12 0 mA 4 0935 4 81 6 F For Voltage output in C 20 to 75 C C Output measured in volts 010526 20 C EXAMPLE 0 50 Vdc 010526 20 27 5 C For Voltage output F 4 to 167 F F Output measured in volts...

Страница 11: ...2 Connect transmitter as shown in figure below 3 Apply power to transmitter and allow to warm up for 15 min 4 Place unit in a 11 RH environment and allow to stabilize for 15 min 5 Adjust potentiomete...

Страница 12: ...main board as shown below 5 Apply power to transmitter and allow to warm up for 10 min 6 Set the RTD simulator to the 0 F position 7 Adjust potentiometer P5 on the main board for a output reading of 4...

Страница 13: ...ace unit in a 11 RH environment and allow to stabilize for 15 min 5 Adjust potentiometer P3 for a output reading of 110 Vdc 6 Place unit in a 75 RH environment and allow to stabilize for 15 min 7 Adju...

Страница 14: ...er on the main board as shown below 5 Apply power to transmitter and allow to warm up for 10 min 6 Set the RTD simulator to the 0 F position 7 Adjust potentiometer P5 on the main board for a output re...

Страница 15: ...93 1 Remove enclosure cover 2 Connect transmitter as shown in figure 6 3 Apply power to transmitter and allow to warm up for 15 min 4 Place unit in a 11 RH environment and allow to stabilize for 15 m...

Страница 16: ...mulator for 150 F output 9 Adjust potentiometer GAIN on the back of the RH display for a reading of 150 0 F 10 Repeat steps 6 7 8 and 9 as necessary until proper readings are maintained 11 Calibration...

Страница 17: ...ensing Accuracy 2 5 72 F with an added temperature coefficient error of 06 RH F both increasing and decreasing in temperature Repeatability 1 RH Operating Temperature Range 20 to 75 C 4 to 167 F Outpu...

Страница 18: ...d up to 13 UL Listed Connections Cable Strain Relief Nylon Liquid tight with neoprene gland for 2 29 to 6 73 mm 09 to 265 diameter cable Wire Connections Internal 3 Position Terminal Strip Accepts fro...

Страница 19: ...e shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEG...

Страница 20: ...tors Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conduc...

Отзывы: