Section 7 - Maintenance

The sensor probe can be replaced quickly for Wall Mount and Remote Probe 

models.
Procedure to replace the sensor probe:
a. Loosen the securing nut, and pull the sensor probe out.
b. I nsert the new sensor probe to the M12 connector, and tighten the securing 

nut to secure the connection.

Figure 7-1  Sensor Probe Replacement

If your HX92B will be used in a dusty environment, the protective sensor filter 

(if clogged) may be removed for cleaning.
Unscrew the protective cover and gently blow compressed air through the filter 

screen. A soft brush may also be used to remove dirt particles from the screen. 
If the sensor is subjected to 100% condensation, it must be dried to obtain correct 

readings. There will be no permanent damage or calibration shift to the unit.
Units should not be exposed to high concentrations of ammonia or alcohol 

vapors.
The Duct Mount model transmitter must be sent back to the factory if sensor 

replacement or recalibration is needed.

SECURING NUT

REPLACEABLE

SENSOR PROBE

HX92B SERIES

RELA

TIVE HUMIDITY

TRANSMITTER

7-1

Maintenance

7

Содержание HX92B Series

Страница 1: ...tor HX92B D duct mount transmitter shown smaller than actual size HX92B wall mount shown larger than actual size TH SP short probe with M12 connector Easy M12 Connection TM User sGuide Shop online at...

Страница 2: ...y U S and International Patents and Patents Pending omega com info omega com Omega Engineering Inc 800 Connecticut Ave Suite 5N01 Norwalk CT 06854 USA Toll Free 1 800 826 6342 USA Canada only Customer...

Страница 3: ...1 1 2 Unpacking 2 1 3 Theory of Operation and Electrical Connection 3 1 3 1 1 Theory Operation 3 1 4 Mounting 4 1 5 Relative Humidity Output Calculations 5 1 6 Calibration 6 1 7 Maintenance 7 1 8 Spec...

Страница 4: ...ter Internal Diagram 1 1 Figure 3 1 Voltage Output Connection Diagram 3 1 Figure 3 2 Current Output Connection Diagram 3 2 Figure 4 1 Wall Mount Model Dimensions 4 1 Figure 4 2 Duct Mount Model Dimens...

Страница 5: ...ight protection Screws are provided for mounting via internal holes inside the enclosure The terminal connection for the output signal is as shown in Figure 1 1 below Figure 1 1 Transmitter Internal D...

Страница 6: ...inspect the container and equipment for any signs of damage Note any evidence of rough handling in transit Immediately report any damage to the shipping agent The carrier will not honor any damage cl...

Страница 7: ...obe Relative Humidity Transmitter 4 to 20mA Output with 3 m 10 cable HX92BV0 RP1 Remote Probe Relative Humidity Transmitter 0 to 1 volt output with 3 m 10 cable HX92BV1 RP1 Remote Probe Relative Humid...

Страница 8: ...a current loop will normally consist of a power supply the transmitter and a meter recorder or controller to measure the current flow The maximum allowable loop resistance for the HX92B to function p...

Страница 9: ...Figure 3 2 Current Output Connection Diagram 3 2 POWER IN R H OUTPUT POWER SUPPLY 9 30 Vdc MAX 4 20 MA GND NC NC TB1 GND CHASSIS 1 2 3 4 5 Theory of Operation and Electrical Connection 3...

Страница 10: ...ded with Wall Mount and Remote Probe models A duct mounting kit is included with Duct Mount models See Figure 4 1 Figure 4 2 and Figure 4 3 for the Wall Mount Duct Mount and Remote Probe model dimensi...

Страница 11: ...Probe Model Dimensions 4 2 17 3 0 68 54 1 2 13 304 8 cm 10 127 3 5 01 65 2 56 50 1 97 DIMENSIONS mm in HX92B SERIES RELATIVE HUMIDITY TRANSMITTER 163 3 6 43 16 0 0 63 65 0 2 56 128 0 5 04 50 0 1 97 3...

Страница 12: ...voltage output use the following formulas For current output RH Current measured in milliamps 4 0 16 Where 4 offset current in mA 0 16 full scale current limit in percent For voltage output a 1 volt...

Страница 13: ...0 75 1 50 20 7 20 2 20 1 00 2 00 25 8 00 0 25 1 25 2 50 30 8 80 0 30 1 50 3 00 35 9 60 0 35 1 75 3 50 40 10 40 0 40 2 00 4 00 45 11 20 0 45 2 25 4 50 50 12 00 0 50 2 50 5 00 55 12 80 0 55 2 75 5 50 6...

Страница 14: ...Calibration Your transmitter has been digitally calibrated and tested to meet or exceed the specifications outlined in this manual in our factory The transmitter must be sent back to the factory for...

Страница 15: ...be removed for cleaning Unscrew the protective cover and gently blow compressed air through the filter screen A soft brush may also be used to remove dirt particles from the screen If the sensor is su...

Страница 16: ...e 200 Ohms 9 Vdc supply voltage 4 to 20mA 1 300 Ohms 30 Vdc supply voltage Ohms V supply 4 V 02 A Max Load min resistance Resistance 1 25K for outputs 0 1 Vdc 0 5 Vdc 2 40K for 0 10 Vdc output Sensor...

Страница 17: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Страница 18: ...s M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U...

Отзывы: