One Analog Output:
4-20mA Loop Power corresponds -60 to 40C dew point
Output signal resolution:
0.03C
Power Supply:
12 to 28Vdc, 25 mA max.
Display resolutions:
0.1C
Dew Point:
Accuracy: +/- 1.0C from 40c to -20C, +/-3.0C <-20C
Temp. Accuracy:
+/- 0.5°C from 0 to 100C, >100C +/-1C
Mounting: Cable Length: 1 meter (except wall units)
SS Material:
Probe, filter guard and fitting
Fitting:
½” NPT stainless steel (included)
Pressure rating:
750psi
Dimensions of Probe:
Length: 6.5” (185mm), O.D.: 0.5” (13mm)
For the HX200-RS series
The HX200-RS meets all the spec
’
s listed above for both the HX200HD and HX200HR.
Digital Output:
Bidirectional RS-232C outputs will display RH, dew point and temperature, in
the streaming command. The loop power will not function if using the RS232.
Switches for DP or RH and Fahrenheit or Celsius
Located on the circuit board there are two switches. The switch on the left will allow
for the customer to switch between DP or RH and the switch on the right allows for
the customer to switch between Celsius and Fahrenheit.