OPERATING INSTRUCTIONS

 

1. OPERATIONAL MODE 

There  are  three  operation  modes-Normal,  Shift,  and 
Setup Mode. 
NORMAL MODE: 

This  is  the  default  mode,  the  operating  functions  for 
the normal mode are printed on the top of each button 
in white. 

SHIFT MODE: 

The operating functions for the shift mode are printed 
in  gray  on  the  buttons.  While  in  the  normal  mode, 
press the SHIFT button to switch to shift mode. At the 
lower-right  corner  of  the  display,  the  word  “Shift”  is 
displayed  to  indicate  shift  mode.  To  switch  back  to 
normal mode, press the SHIFT button again. 

SETUP MODE: 

Press  the  set[  ]  button  in  normal  mode  to  switch  to 
setup  mode,  the  indicator  “SET”  is  shown  on  the  left 
side  of  the  display.  To  switch  back  to  normal  mode, 
press SET[ ] button. 
 

2. NORMAL MODE 

The  following  functions  can  only  be  used  in  the  normal 
mode. 

(1) “

” Power Button 

The  “

”  button  turns  the  thermometer  on  or  off. 

When  the  meter  is  in  MAX/MIN  record  mode,  the 
power off function is disabled. 

(2) “[LIMITS]” Button (only Main display) 

The  limits  function  will  alert  the  user  when  a  meas-
urement exceeds a specified limit. To set the limit val-
ues,  refer  to  limits  function  in  the  setup  mode.  Press 
the [LIMITS] button to activate the limits function; the 
word “LIMIT” should be displayed on the LCD. 
When  the  value  of  the  main  display  exceeds  the  Hi 
limit,  the  word  “Hi”  will  be  displayed  and  the  ther-
mometer will beep in a pulsed tone. If the value of the 
main display is lower then the Lo limit, the word “Lo” 
will  be  displayed  and  the  thermometer  will  beep  in  a 
continuous  tone.  To  exit  the  limits  function,  press  the 
[LIMITS] button. 

(3) “ ” Button 

The  backlight  function  is  represented  by  this  button 

”.  Pressing  the  button  will  turn  on or  off  the  LCD 

backlight. The backlight will turn off automatically af-
ter. 

(4) “SAVE/READ” Button 

The  read  data  function  works  in  conjunction  with  the 
save  function  in  the  shift  mode,  it  is  used  for  reading 
saved data. The save function can be activated in shift 
mode.  Press  the  SAVE/READ  button  to  read  saved 
data;  the  word  “READ”  should  be  displayed  on  the 
LCD. To navigate around the save data table, press the 
overlay  “SECOND”  button  until  the  “#”  sign  is  dis-
played on the second display. The location of the read 
pointer  within  the  saved  data  table  will  be  displayed. 
The arrow buttons on the overlay are used for scrolling 
through  the  saved  data.  Press  the  smaller  arrows  “

” 

or “

” to step through the data one at a time. Press the 

larger  arrows  “

”  or  “

”  to  step  through  the  data 

ten at a time. Pressing the overlay “ESC” button deac-
tivates the read data function. 

(5) “LOG/READ” Button 

The  read  log  function  works  in  conjunction  with  the 
log function. It is used for reading logged data. The log 
function  can  be  activated  in  the  shift  mode.  Press  the 
LOG/READ  button  to  activate  the  log  read  function; 
the word “READ” is displayed on the  LCD. Press the 
overlay  SECOND  button  to  rotate  through  following 
display  menus:  T1,  T2,  GRP,  and  #.  T1  and  T2:  Dis-
plays the T1 or T2 saved data. GRP: Displays the cur-
rent group number. #: Displays the current location of 
the  read  pointer  within  a  selected  group.  The  arrow 
buttons  on  the  overlay  are  used  for  scrolling  through 
the  data  and  groups.  Press  the  smaller  arrows  “

”  or 

” to step through the logged data or groups one at a 

time.  Press  the  larger  arrows  “

”  or  “

”  to  step 

through  the  data  or  groups  ten  at  a  time.  To  navigate 
the  logged  data  and  groups,  press  the  overlay 
SECOND button until GRP appears in the second dis-
play  panel.  Then  select  the  group  using  the  arrows. 
Press  the  SECOND  button  again  until  the  “#”  sign  is 
displayed.  The  location  of  the  read  pointer  in  the  se-
lected  group  will  be  displayed.    Use  the  arrows  to 
scroll  through  the  data.  Pressing  the  overlay  “ESC” 
button deactivates the read data function.

   

(7) HOLD Mode (only Main display) 

When HOLD mode is selected, the thermometer holds 
the  present  readings  and  stops  all  further  measure-
ments.  To  activate  the  data  hold  mode,  press  the 
HOLD button, and “HOLD” is displayed on the LCD. 
Pressing the HOLD button again cancels the function, 
and  the  instrument  will  automatically  resume  meas-
urements. 

(8) MIN/MAX with Time record Mode 

(only Main display) 

The  MIN/MAX  function  records  the highest  and  low-
est  value  recorded,  and  it  calculates  the  average  read-
ing,  and  the  differences  of  MAX  to  MIN.  Press 
MIN/MAX  button  to  enter  the  MIN/MAX  recording 
mode. The beeper emits a tone when a new minimum 
or  maximum  measurement  is  recorded.  Press  the 
MIN/MAX  button  again  to  rotate  through  the  current 
readings:  MAX:  The  highest  measurement  recorded. 
MIN:  The  lowest  measurement  recorded.  MAX-MIN: 
The difference of the highest and the lowest measure-
ment. AVG: The average values of the measurements. 
This mode works in conjunction with the hold function, 
pressing the HOLD button will stop the recording and 
measurements  (Previously  recorded  readings  are  not 
erased). Press HOLD button again to resume recording 
and measurements. To prevent accidental loss of MIN, 
MAX and AVG data, the MIN/MAX function can only 
be  cancelled  by  pressing  and  holding  down  the  MIN 
MAX  key  for  more  then  2  seconds.  The  automatic 
power-off  feature,  and  the  power,  °C/°F,  REL,  SET, 
Hi/Lo Limits, TYPE, T1/T2 buttons are also disabled. 

(10) “T1/T2” Button (Main display) 

The input selection button [T1/T2] selects the input for 
the  main  display,  T1  thermocouple  or  T2  thermocou-
ple. Press the T1/T2 button to switch between the two 
inputs. When meter is turned on, it is set to the display 
that was last in use. 

(11) “T1/T2” Button (Second display) 

The input selection button [T1/T2] selects the input for 
the  second  display,  T1  thermocouple  or  T2  thermo-
couple.  Press  the  T1/T2  button  to  switch  between  the 
two  inputs.  When  meter  is  turned  on,  it  is  set  to  the 
display that was last in use. 

(12) “T1-T2/TIME” Button (Third display) 

The  input  selection  button  [T1-T2]  selects  the  system 
time and date, or the differential between the two ther-
mocouples  (T1-T2)  for  the  third  display.  Press  the 
T1-T2 button to switch the display options. When me-
ter is turned on, it is set to the display that was last in 
use. 

 

3. SHIFT MODE 

The  following  functions  can  only  be  used  in  the  shift 
mode. 

(2) “°C/°F” Button 

Press the °C/°F button to select the temperature scale, 
readings can be displayed in Celsius (°C) or Fahrenheit 
(°F).  When  the  thermometer  is  turned  on,  it  is  set  to 
the temperature scale that was last in use. 

(4) “SAVE” Button 

The save function stores the T1, T2 data in a nonvola-
tile  memory.  Press  the  SAVE  button  to  save  the  cur-
rent  data,  the  word  SAVE  is  displayed  to  indicate  the 
data has been saved. The built in memory can store up 
to 128 data (HH806U)/256 data (HH806AU). The data 
can  be  read  using  the  read  function  in  the  normal 
mode. 

(5) “LOG” Button (HH806U) 

The  data  log  function  continuously  records  the  data 
according  to  a  specified  time  interval.  The  time  inter-
val can be set using the interval setup function [INVT] 
in  the  set  up  mode.  Press  the  LOG  button  to  activate 
the  log  function;  the  indicators  “LOG”  and  “MEM” 
will be displayed on the LCD. There are 16 groups that 
are  used  for  storing  the  log  data, and  each  group uses 
64 data slots. If the current log session exceeds 64 data, 
the  log  function  will  automatically  use  the next  group 
to  store  the  following  data.  A  maximum  of  1024  data 
point  can be  saved  in  one  log  session.  Press  the  LOG 
button again to exit the data log function.

   

“LOG” Button (HH806AU) 

The  data  log  function  continuously  records  the  data 
according  to  a  specified  time  interval.  The  time  inter-
val can be set using the interval setup function [INVT] 
in  the  set  up  mode.  Press  the  LOG  button  to  activate 
the  log  function;  the  indicators  “LOG”  and  “MEM” 
will be displayed on the LCD. There are 16 groups that 
are  used  for  storing  the  log  data.  A  maximum  of 
16,000  data  point  can  be  saved  in  one  log  session. 
Pressing  the  LOG  button  again  to  exit  the  data  log 

function. 

(6) “CLR ?” Button 

The CLR function clears all the saved and logged data 
in memory. When the CLR button is pressed, indicator 
“MEM” is displayed and the “CLR” on upper-right of 
LCD will blink. Pressing the “ENTER” button printed 
on the overlay in white will clear all saved and logged 
data.  Press  “ESC”  button  to  exit this  function  without 
clearing data. 

(7) “REL” Button (Main display) 

The relative value function can be used for comparing 
the  saved  reference  value  with  other  measurements. 
Press  the  “REL”  button  to  store  the  current  measure-
ment as the reference value, and “REL” should be dis-
played  on  the  right  part  of  the  LCD.  The  next  meas-
urement will display the current value compared to the 
reference value. Press “REL” button again to clear the 
reference value and deactivate the relative value meas-
urement function. 

(8) “APO” Button 

Press  the  APO  button  to  turn  the  “Auto  power  off” 
function  on  or  off.  When  this  function  is  enabled,  the 
indicator “APO” is shown at the upper left part of the 
LCD.  When  APO  (Auto  power  off)  is  enabled, it  will 
automatically  turn  the  thermometer  off  no  button  is 
pressed  for  a  period  longer  than  the  set  time  interval 
(the  default  time  for  APO  is  5  minutes).  Press  power 
button to resume operation. 

(10) “TYPE” Button (Main display) 

Press this button to change the type of thermocouple in 
the  main  display  (K/J/T/E/R/S/N).  If  the  inputs  of  the 
main  and  second  display  are  the  same,  then  pressing 
this button will change the thermocouple type for both 
displays. 

(11) “TYPE” Button (Second display) 

Press this button to change the type of thermocouple in 
the second display (K/J/T/E/R/S/N). If the inputs of the 
main  and  second  display  are  the  same,  then  pressing 
this button will change the thermocouple type for both 
displays. 
 

4. SETUP MODE 

The  following  functions  can  only  be  use  in  the  setup 
mode. 

(2) “[LIMITS]” Button (Hi/Lo limit setting) 

Press the LIMITS button to enter the Hi/Lo limit setup 
function.  The  words  “LIMIT”  and  “Hi”  will  be  flash 
on  the  LCD  along  with  the  previous  value  for  the  Hi 
limit.  Enter  the  new  Hi  limit  value  using  the  number 
keys printed in white on the overlay. The resolution of 
Hi/Lo  limit  setup  is  1  count.  The  “-”  button  (same 
button  as  the  ESC)  can  be  used  to  enter  negative  val-
ues.    Press  the  “ENTER”  button  to  confirm  the  new 
limit.    You  will  then  be  prompted  to  enter  a  new  Lo 
limit value.    Enter the new value and then press enter 
to finish setup of limits. 

(3) “[INVT]” Button (Interval time setting) 

To  setup  the  time  interval  for  the  log  function,  press 
the [INVT]  button.  The  indicator  “INV”  will  blink  on 