background image

OMEGA

HH501AS/HH501BS/HH501AR/HH501BR

Digital Thermometer

Содержание HH501AR

Страница 1: ...OMEGA HH501AS HH501BS HH501AR HH501BR Digital Thermometer...

Страница 2: ...l Free in Germany 0800 82 66342 e mail nl omega com e mail germany omega com Czech Republic United Kingdom ISO 9002 Certified ul Rude armady 1868 733 01 Karvina One Omega Drive Hranice Czech Repubic R...

Страница 3: ...CAUTION Repeated sharp flexing can break the thermocouple leads To prolong lead life avoid sharp bends in the leads especially near the connector INTRODUCTION This instrument is a 3 digit compact size...

Страница 4: ...ns 195mm H x 92mm W x 53mm D Weight 365g Input Connector Accepts standard miniature thermocouple connectors flat blades spaced 7 9mm center to center Battery Life 100 hours typical battery Auto power...

Страница 5: ...us or Fahrenheit user selectable Measurement Range 0 C to 1767 C 32 F to 1999 F Resolution 1 C 1 F Accuracy Accuracy is specified for operating temperatures over the range of 18 C to 28 C 64 F to 82 F...

Страница 6: ...the temperature scale press the C or F key Button Press button to toggle on and off of backlight The backlight will switch off automatically after 30 seconds HOLD Mode Pressing the HOLD key to enter t...

Страница 7: ...of readings all values are frozen press again to restart recording To prevent accidental loss of MIN MAX MAX MIN and AVG data in this mode can only be cancelled by pressing and hold down the MIN MAX...

Страница 8: ...Relative mode zero the display and store the displayed Reading as a reference value and annunciator REL is displayed Press REL key again to exit the relative mode The relative value can also be entere...

Страница 9: ...scape Hi Limit value set mode _ is displayed the is displayed when you set up the fourth number Hi Limit value is entered via overlay numbers when you want to get negative values push 0 key for end of...

Страница 10: ...ic power off feature is disabled and other funtion key are disabled when input temperature value exceed Hi The beeper emits a continuity pulse tone and Hi is displayed and when input temperature value...

Страница 11: ...eter before removing the cover Battery Replacement Power is supplied by a 9 volt transistor battery NEDA 1604 IEC 6F22 The appears on the LCD display when replacement is needed To replace the battery...

Страница 12: ...on J1 position 2 Input DCV 25 6mV to the T1 after the display is stabilized forward 4 digits then press ENTER key and 01 is displayed 3 Input DCV 74mV after the display is stabilized then press ENTER...

Страница 13: ...ty or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipmen...

Страница 14: ...Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivi...

Отзывы: