DFG35 Digital Force Gauges
User’s Guide
12
The gauge can be calibrated at up to 10 points in each direction. Enter the number of calibration
points for each direction (compression and tension). At least one point must be selected for each
direction.
Note:
To achieve the accuracy specification of ±0.3%, it is recommended to calibrate the gauge
at 5 or more even increments in both the tension and compression directions. For example, a
gauge with capacity of 10 lbF should be calibrated at 2, 4, 6, 8, and 10 lb loads in each direction.
3. To escape the
Calibration
menu at any time, press
ESCAPE
. The display will appear as follows:
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will
return to the menu without saving changes.
4. After the number of calibration points has been entered, press
ENTER
. The display will appear as
follows:
5. Place the force gauge horizontally on a level surface free from vibration, then press
ZERO
. The
gauge will calculate offsets, and the display will appear as follows:
If failed:
6. The following screen appears after the offsets have been calculated:
CALIBRATION
NOT COMPLETE
Cancel
Exit w/o saving
CALIBRATION
OFFSET
Sensor failed
Analog failed
CALIBRATION
OFFSET
Sensor passed
Analog passed
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place force gauge
horizontally, then
press ZERO.