36
4.5.1 HTTPget using Port 1000 (continued)
httpget -r -S “*SRTC\r” 192.168.1.200:2000
where:
-r –S
are parameters needed for the command string
*SRTC
is the command as the following:
*SRTC
Read the temperature in C (Channel 1)
*SRTF
Read the temperature in F (Channel 1)
*SRHC
Read temperature in C (Channel 2)
*SRHF
Read temperature in F (Channel 2)
*SRDC
Read the differential in C
*SRDF
Read the differential in F
*SRA1
Read temperature setpoints (Channel 1)
*SRA2
Read temperature setpoints (Channel 2)
*SRB
Read the temperature in C, (both Channel 1 and Channel 2)
*SRYS
Indicate AC or DC power and battery status
*SRYRST
Reset Power on iSE
*SR_XD1
Exclude alarm 1 (disable alarm 1 function)
*SR_XD2
Exclude alarm 2 (disable alarm 2 function)
*SR_XC1
Reset relay 1 latch condition
*SR_XC2
Reset relay 2 latch condition
*SR_XE1
Include alarm 1 (enable alarm 1 function)
*SR_XE2
Include alarm 2 (enable alarm 2 function)
*SR_XP1
Close relay 1 for < 200 mSec
*SR_XP2
Close relay 2 for < 200 mSec
*SR_XH1
Close relay 1
*SR_XH2
Close relay 2
*SR_XL1
Open relay 1
*SR_XL2
Open relay 2
*SRT2
Status of Contact 1 (Disable, Normal, Active)
*SRH2
Status of Contact 2 (Disable, Normal, Active)
*SRD2
Status of the Output associated with the Contacts (Active High,
Active Low)
\r
is the carriage return termination character
192.168.1.200
is an IP address
2000
is a socket port number
Response:
076.6 (in Deg.F 35)
Содержание aiSE-TC
Страница 57: ...NOTES...