3)  Raise the condenser till the top lens is close to the opening of stage. Place a drop of 

immersion oil on the top of the lens of the condenser. 

4)  Place the slide on the stage. Raise the condenser and let the oil 

drop contact the bottom of the slide. If air bubbles exist in the oil, 
clean the oil from the condenser lens and bottom of slide with a 
camera cleaning paper and repeat the procedures.   

5)  Follow the procedures in this manual to bring the sample on slide 

in focus and observe. 

6)  When using the 100X oil darkfield objective, you will need to drop 

the immersion oil and adjust the iris ring (

Fig. 14

) to get proper 

brightness and contrast of view field. 

Note

 

 

The condenser won

’t work well if no oil drop applied on the condenser

 

 

 
 

8.

 

Phase Contrast Kit Installation and Operation 
Instructions 

8.1  Phase contrast kits of A1PHB1 or A1PHB3: 

1)  Mounting the phase contrast objectives 

a. Take off all the objectives from the nosepiece. 
b. Install the phase contrast objectives onto the nosepiece following the steps in 

2.3

2)  Mounting the phase contrast condenser/annular ring disk 

a. Rotate the nosepiece and set the 4X objective in position.

 

b. Turn the coarse focus knob to raise the mechanical stage to the highest position 

without contacting the 4X objective.

 

c.  Turn the condenser control knob to raise the condenser to the highest position.

 

d. Loosen the light collector and take it off.

 

e. Loosen the condenser lock thumb screw, pull down 

the condenser and take it off.

 

f.  Insert the condenser/annular ring disk into the 

condenser holder, and tighten the thumb screw.

 

g. Insert the light collector back onto the base and turn 

until it’s tightened.

 

h. Re-adjust the condenser height to its normal 

position.

 

i.  Re-adjust the mechanical stage height to its normal 

position.

 

Note:   

  When raising the mechanical stage, do not make 

contact with the objective. 

3)  Turn the desired objective into light path. 
4)  Turn the annular ring disk to put the corresponding 

ring into light path, i.e. if you are using the 

20X

 

phase contrast objective, you should turn the disk at 

Fig.14 

  Iris Ring 

20 

Fig.15 

Fig.16 

Содержание M8311Z Series

Страница 1: ...User Manual Trinocular Compound LED Microscope Model M8311Z Series MicroscopeNet com...

Страница 2: ...1 Components Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 9 5 Optional Parts 10 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Страница 3: ...he container should be retained should the microscope ever requires reshipment 3 Keep the instrument out of direct sunlight high temperature or humidity and dusty environments Ensure that the microsco...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...al 3 Diopter Ring 11 Nosepiece 19 Condenser Lock Thumb Screw 4 Eyepiece Tube 12 Objective 20 Condenser Focus Knob 5 Viewing Head 13 Microscope Body 21 Condenser 6 Slide Holder 14 Coarse Focus Knob 22...

Страница 6: ...s from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn t...

Страница 7: ...on the photo tube 2 Insert the camera into the photo tube and then connect the camera to a computer via USB2 0 cable See Fig 4 3 The manual for the camera is on a CD or mini CD Refer to the manual to...

Страница 8: ...imen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus Then use the fine focus knob to obtain a sharp image You may now switch to another magnification objective 3...

Страница 9: ...objectives When you use the 100X objective to observe the specimen you have to apply the immersion oil on the top of the slide cover When you apply the immersion oil with the 100X objective do not le...

Страница 10: ...procedures in 3 3 2 Install the camera by following the procedures in 2 7 3 Open image observing software to examine more details see camera s manual 4 If the image is not clear loosen the 2 set scre...

Страница 11: ...lter holder Rack and pinion focus adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with stop Stage Double layer mechanical stage Dimension 5 1 2 x 4 1 4 140mm x 110mm Tr...

Страница 12: ...Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above A3520U 1600 x 1200 2 0MP A3530U 2048 x 1536 3 0MP A3550U 2592 x 1944 5 0MP A3590U 3488x2616 9 0MP...

Страница 13: ...eter A191 Dry 0 7 0 9 37mm A191BD Oil 1 36 1 25 37mm A191BOIL new design Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring with iris diaphragm 37mm 3 Carrying Case Model Dimension Net Weight A187V...

Страница 14: ...Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for 20X phase contrast objective 40 for 40X...

Страница 15: ...the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cove...

Страница 16: ...nsufficient or too bright The intensity of lamp is too low Adjust the intensity dial to increase the brightness The condenser is not in the right position Lower or raise the condenser slightly to the...

Страница 17: ...erve 6 Move the condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X...

Страница 18: ...piece b Install the phase contrast objectives onto the nosepiece following the steps in 2 3 2 Mounting the phase contrast condenser annular ring disk a Rotate the nosepiece and set the 4X objective in...

Страница 19: ...nular ring disk being put at B position 6 Performing the phase contrast observation After centering the condenser ring plate perform the phase contrast observation the same way as a normal bright fiel...

Страница 20: ...rk ring should be coincided with each other as shown in Fig 19 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f If the bright ring is still obscure...

Страница 21: ...bjective do not let the immersion oil to contact with the dry objective lenses especially the 40X If the immersion oil is on the dry objectives lens please use the lens cleaning paper to clean the obj...

Страница 22: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 23 a h If the bright ring is still obscure as in Fig 23 b raise or l...

Отзывы: