Darkfield Condenser Installation and 
Operation Instructions 

7.1 

Dry darkfield condenser of A191 

1)  Mounting the dry darkfield condenser 

  Loosen the condenser lock thumb screw on the 

condenser holder and remove the brightfield 
condenser. 

  Install the dry darkfield condenser and tighten 

the condenser lock thumb screw on the 
condenser holder (see 

Fig. 10

)

2)  Centering the dry darkfield condenser 

  Turn the 4X objective to the light path.   

  Turn the condenser focus knob to lower the 

condenser till a dark spot showed in the viewing 
field as shown in 

Fig. 11

 

(a)

  Turn the condenser translational centering screws 

to move the dark spot to the center as shown in 

Fig. 11

 

(b)

3)  Place the slide on the stage.   
4)  Raise the condenser all the way to the top and then 

lower it a little bit.   

5)  Follow the procedures in this manual to bring the sample on slide in focus and 

observe. 

6)  Move the condenser up or down slightly to get the best darkfield viewing. 

Note

 

  The dry darkfield condenser is used with the dry objectives only. 

  The dry darkfield condenser works with the 4X, 10X, 40X objectives. 

 

The dry darkfield condenser does not work with the 100X oil immersion objective.

 

7.2  Oil darkfield condenser of A191BOIL 

1)  Mounting the oil darkfield condensers 

  Loosen the condenser lock thumb screw on the 

condenser holder and remove the brightfield 
condenser. 

  Install the oil darkfield condensers and tighten 

the condenser lock thumb screw on the 
condenser holder (see

 

Fig. 12

)

2)  Centering the oil darkfield condensers 

  Turn the 40X objective to the light path.   

  Turn the condenser focus knob and slowly lower 

or raise the condenser till a dark spot showed in 
the viewing field as shown in 

Fig. 13

 

(a)

  Turn the condenser translational centering 

screws to move the dark spot to the center as 
shown in 

Fig. 13 (b)

 

     

(a)

                             

  (b)

 

 

   

 

   

(a)

                               

(b) 

Fig.10 

Fig.12 

Fig.13 

Fig.11 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

Содержание M8211 Series

Страница 1: ...User Manual Trinocular Compound LED Microscope Model M8311 Series MicroscopeNet com...

Страница 2: ...1 Components Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 9 5 Optional Parts 10 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Страница 3: ...he container should be retained should the microscope ever requires reshipment 3 Keep the instrument out of direct sunlight high temperature or humidity and dusty environments Ensure that the microsco...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...al 3 Diopter Ring 11 Nosepiece 19 Condenser Lock Thumb Screw 4 Eyepiece Tube 12 Objective 20 Condenser Focus Knob 5 Viewing Head 13 Microscope Body 21 Condenser 6 Slide Holder 14 Coarse Focus Knob 22...

Страница 6: ...s from the eyepiece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn t...

Страница 7: ...on the photo tube 2 Insert the camera into the photo tube and then connect the camera to a computer via USB2 0 cable See Fig 4 3 The manual for the camera is on a CD or mini CD Refer to the manual to...

Страница 8: ...imen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus Then use the fine focus knob to obtain a sharp image You may now switch to another magnification objective 3...

Страница 9: ...objectives When you use the 100X objective to observe the specimen you have to apply the immersion oil on the top of the slide cover When you apply the immersion oil with the 100X objective do not le...

Страница 10: ...procedures in 3 3 2 Install the camera by following the procedures in 2 7 3 Open image observing software to examine more details see camera s manual 4 If the image is not clear loosen the 2 set scre...

Страница 11: ...d filter holder Rack and pinion focus adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with stop Stage Double layer mechanical stage Dimension 5 1 2 x 4 1 4 140mm x 110m...

Страница 12: ...Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above A3520U 1600 x 1200 2 0MP A3530U 2048 x 1536 3 0MP A3550U 2592 x 1944 5 0MP A3590U 3488x2616 9 0MP...

Страница 13: ...eter A191 Dry 0 7 0 9 37mm A191BD Oil 1 36 1 25 37mm A191BOIL new design Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring with iris diaphragm 37mm 3 Carrying Case Model Dimension Net Weight A187V...

Страница 14: ...Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil Ffive positions 10 for 10X phase contrast objective 20 for 20X phase contrast objective 40 for 40X...

Страница 15: ...the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cove...

Страница 16: ...nsufficient or too bright The intensity of lamp is too low Adjust the intensity dial to increase the brightness The condenser is not in the right position Lower or raise the condenser slightly to the...

Страница 17: ...erve 6 Move the condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X...

Страница 18: ...piece b Install the phase contrast objectives onto the nosepiece following the steps in 2 3 2 Mounting the phase contrast condenser annular ring disk a Rotate the nosepiece and set the 4X objective in...

Страница 19: ...nular ring disk being put at B position 6 Performing the phase contrast observation After centering the condenser ring plate perform the phase contrast observation the same way as a normal bright fiel...

Страница 20: ...rk ring should be coincided with each other as shown in Fig 19 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f If the bright ring is still obscure...

Страница 21: ...bjective do not let the immersion oil to contact with the dry objective lenses especially the 40X If the immersion oil is on the dry objectives lens please use the lens cleaning paper to clean the obj...

Страница 22: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 23 a h If the bright ring is still obscure as in Fig 23 b raise or l...

Отзывы: