Important Information — Please Read Before Use
UCR INSTRUCTION MANUAL
Use in combination with the lens cleaning sheath
WARNING
• Anytime you observe an irregularity in the endoscopic CO
2
regulation unit,
immediately stop using the endoscopic CO
2
regulation unit and withdraw the
endoscope from the patient slowly as described in the instruction manual for the
lens cleaning sheath.
• To prevent gas embolism caused by intra-abdominal over-pressurization from the
combined use of the lens cleaning sheath and a laser device, argon-enhanced
coagulator, or other gas supply devices, carefully read and understand the
following before using in case the insufflator is used in operation:
−
When performing endoscopic operation using a lens cleaning sheath and a
laser device, argon-enhanced coagulator, or other gas supply devices
simultaneously, both instruments become a source of supply for gas.
Accordingly, the desired pressure in the abdomen is reached in a shorter time
than when a lens cleaning sheath is used alone. In these cases, be careful that
the cavity does not become over-pressurized. The laser device,
argon-enhanced coagulator, and other gas supply devices are not equipped to
monitor the cavity pressure (automatic termination of insufflation, warning light
or alarm). Although the insufflator is equipped with these functions, it does not
always prevent gas embolism inasmuch as this depends on the patient and the
condition of the infected area. We ask that the physicians make a suitable
judgment from a professional standpoint himself.
−
If the insufflator emits a caution (caution light or alarm) for cavity
over-pressurization, quickly open the stopcock or valve of the trocar. Then,
reduce the amount of outflow from the laser device, argon-enhanced
coagulator, or other gas supply device. If use is continued while the alarm
sounds, there is a risk of gas embolism due to cavity over-pressurization.
• Be careful not to feed CO
2
too much during CO
2
gas feed or spraying. Excessive
CO
2
gas feed/spraying may cause pain to the patient and/or gas embolism.
Reduce the use of CO
2
gas feed and spraying at the requisite minimum level.
• In case the insufflator is used in operation, open the trocar cannula stopcock if the
pressure in the abdomen exceeds the set pressure by more than 5 mmHg.
• Do not attempt to insufflate the abdominal cavity using the lens cleaning sheath.
Otherwise, patient injury and/or gas embolism may result.
• Do not use the endoscopic CO
2
regulation unit in combination with other than the
compatible lens cleaning sheath (with the MAJ-1345, MAJ-1537, or MAJ-659, for
example). Otherwise, CO
2
gas is fed continuously and patient pain and gas
embolism can result.
Содержание UCR
Страница 2: ......
Страница 5: ...Contents iii UCR INSTRUCTION MANUAL Specifications 69 EMC information 73...
Страница 6: ...iv Contents UCR INSTRUCTION MANUAL...
Страница 24: ...18 2 2 Front panel UCR INSTRUCTION MANUAL Ch 2 2 2 Front panel 3 1 2 4 5 6 7...
Страница 26: ...20 2 3 Rear panel UCR INSTRUCTION MANUAL Ch 2 2 3 Rear panel 1 2 3 4 5 6...
Страница 50: ...44 4 6 Inspection of the endoscopic CO2 regulation unit UCR INSTRUCTION MANUAL Ch 4...
Страница 58: ...52 5 5 After use UCR INSTRUCTION MANUAL Ch 5...
Страница 70: ...64 6 8 Transportation UCR INSTRUCTION MANUAL Ch 6...