2-2 Mounting the Mercury Burner

High voltage is applied to ignite the mercury burner, to prevent any hazards, conform that the main switch on
the power supply unit is switched to "O" (OFF), unplug the lamp housing cord and power cord plug, and wait
at least 10 minutes before proceeding with the replacement.

Fig. 1

Fig. 2

1. Using the Allen screwdriver provided with the microscope frame, loosen

the burner socket clamping screw @. (Fig. 1)

2. Loosen the burner clamping screws ³ and | and remove the transport

post. (For burner replacement, remove the used burner.) (Fig. 2)

3. Insert the + pole of the mercury burner ² into the + terminal and tighten

the + clamping screw ³.

Then loosen the – clamping screw | (marked UP). Insert the – pole of
the burner into the – terminal and tighten the – clamping screw |. (Fig. 2)

Be sure to use a USH102D burner (mfd. by Ushio Electric) or
HBO103W/2 (mfd. by OSRAM).

Be careful and avoid leaving fingerprints or dirt on the mercury
burner. Attached stain may cause distortion in glass which could
result in a ruptured burner. If the burner is contaminated, clean it
by wiping gently with gauze slightly moistened with absolute
alcohol.

#

To prevent possible damage to the burner, the collector lens can
only be installed or removed while the socket and lamp housing
are separated.

When using the lamp housing U-ULH

Fig. 3

Fig. 4

When using the 100W lamp housing U-LH100HG/LH100HGAPO

1. Loosen the burner socket clamping screw @ using the Allen screwdriver.

(Fig. 3)

2. While holding the upper part of the lamp housing, pull it vertically to

remove the socket.

# Do not hold the centering knob ². Doing so may use malfunction.

3. Place the socket upside down as shown in Fig. 4.

} Loosen the two burner clamping screws ³ to remove the transport lock

when using the unit for the first time or to remove the old burner when
replacing with a new one. (Fig. 4)

4. Securely attach the + pole of the specified mercury burner | to the upper

mount and the – pole to the lower mount. (Fig. 4)

Be sure to use a USH102D burner (mfd. by Ushio Electric) or
HBO103W/2 (mfd. by OSRAM).

Be careful and avoid leaving fingerprints or dirt on the mercury
burner. Attached stain may cause distortion in glass which could
result in a ruptured burner. If the burner is contaminated, clean it
by wiping gently with gauze slightly moistened with absolute
alcohol.

5. Replace the socket with the burner mounted and tighten the burner

socket clamping screw @.

} While aligning the exterior of the lamp housing with that of the socket, let

down the lamp housing vertically until it fits over the socket.

} For details on mounting the collector lens unit and the lamp housing on

the microscope frame, refer to the instruction manuals for the reflected
light fluorescence modules for the AX, BX, BX2, IX, etc., microscope
frames, respectively.

2

Содержание U-RFL-T

Страница 1: ...tain optimum performance and to familiarize yourself fully with the use of this power supply unit we recommend that you also study the instruction manual for the microscope to be used Retain this inst...

Страница 2: ......

Страница 3: ...y sometimes rupture 4 When disposing of the mercury burner always follow national and local laws and guidelines 5 The power supply unit contains high voltage components Never attempt to modify or disa...

Страница 4: ...the equipment Indicates that failure to follow the instructions could result in damage to equipment Indicates commentary for ease of operation and maintenance 1 Getting Ready 1 A power supply unit is...

Страница 5: ...gram 1 2 2 Mounting the Mercury Burner 2 2 3 Cable and Cord Connections 3 3 PREPARATION FOR OBSERVATION 4 3 1 Igniting the Mercury Burner 4 3 2 Centering the Mercury Burner 4 4 SPECIFICATIONS 5 5 TROU...

Страница 6: ...sembly Diagram Lamp housing U ULH Mercury burner configuration Mercury burner USH102D HBO103W 2 Socket U ULS100HG Power Supply Unit U RFL T To illuminator of microscope frame AX BX BX2 IX etc Collecto...

Страница 7: ...using the lamp housing U ULH Fig 3 Fig 4 When using the 100W lamp housing U LH100HG LH100HGAPO 1 Loosen the burner socket clamping screw using the Allen screwdriver Fig 3 2 While holding the upper par...

Страница 8: ...erminal of the power supply unit and that of the wall outlet are properly connected If the equipment is not grounded Olympus can no longer warrant the electrical safety and performance of the equipmen...

Страница 9: ...ATION Fig 9 After turning the burner off it cannot be re ignited before the mercury vapor cools and condenses to a liquid Wait for at least 3 minutes before restarting the burner Opening the lamp hous...

Страница 10: ...87 in H 5 Weight 4 8 kg 10 6 lb Operating Environment Indoor use Altitude Max 2000 m Ambient temperature 5 to 40 C 41 to 104 F Maximum relative humidity 80 for temperatures up to 31 C 88 F decreasing...

Страница 11: ...O OFF Turn to I ON again Repeat as necessary c The burner flickers or is dark Insufficient time has elapsed since Wait for 10 minutes after turning on the the burner was turned on burner The burner l...

Страница 12: ...Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The fit...

Страница 13: ...rds Authority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Electrotecnica Y Electronica Espanola AEE Hellenic Organization for Standardization ELOT Instituto Portages da Qualid...

Страница 14: ...MEMO...

Страница 15: ......

Страница 16: ...rate Center Drive Melville NY 11747 3157 U S A 491B River Valley Road 12 01 04 Valley Point O ce Tower Singapore 248373 2 8 Honduras Street London EC1Y OTX United Kingdom 31 Gilby Road Mt Waverley VIC...

Отзывы: