Attaching the UIS Mirror Unit

(Figs. 2 & 3)

1. Using the Allen screwdriver provided with the microscope frame, loosen

the clamping screw of the mirror unit replacement cover @ on the side
you want to attach the mirror unit.

2. Slide the light path selector lever ² to expose the mount dovetail ³.
3. Remove the light path selector lever ² by turning it in the counter-

clockwise direction, screw it into the threaded screw hole | on the
mount dovetail, and slide out the mount dovetail to expose the mirror
unit mount dovetail ƒ.

4. Slightly loosen the mirror unit clamping screw … using the Allen screw-

driver, fit the dovetail on the mirror unit † all the way into the mount
dovetail ƒ and tighten the clamping screw … firmly.

5. Return the mount dovetail ³ to the original position and remove the

light path selector lever ².

6. Attach the replacement cover @ and light path selector lever ² to the

original positions.

}Perform the same procedure when attaching a UIS mirror unit on the

other side of the U-DPCAD.

@

²

³

|

ƒ

Содержание U-DPCAD

Страница 1: ...l U DPCAD To ensure the safety and obtain optimum performance and to familiarize yourself full with the use of this equipment we recommend that you study this manual thoroughly before operating it Retain this instruction manual in an easily accessible place near the work desk for future reference A X 6 3 0 1 ...

Страница 2: ......

Страница 3: ...scope it will be possible to mount two video cameras on a single microscope The magnification of this adapter is 1X 1 Nomenclature Transmitted light port C mount adapter Detachable Light path selector lever Transmitted light port Reflected light port Mount dovetail Reflected light port Mirror unit replacement cover Provided on both sides Transmitted light port Reflected light port C mount adapter ...

Страница 4: ...ideo camera attached to the reflected light port should not exceed 70 mm x 70 mm Recommended video cameras C2400 77A or C2400 79 by Hamamatsu Photonics CS 220 by Sony etc 4 When a fluorescence mirror unit is attached ghosts or double images may be produced due to the dichroic mirror As the ghosts are particularly noticeable under high intensity in brightfield observation do not use a fluorescence ...

Страница 5: ...Trinocular Observation Tube C mount video camera Side port of IX2 series microscope 2 3 inch CCD video camera FN 11 1 2 inch CCD video camera FN 8 IX2 Side Port Double port adapter U DPCAD C mount video camera Clamping screw UIS trinocular tube C mount video camera C mount video camera Double port adapter U DPCAD Clamping screw ...

Страница 6: ...n the two clamping screws of the C mount adapter for the transmitted light port using the Allen screwdriver then remove the C mount adapter The C mount adapter for the transmitted light port has an identification groove on the bottom side 2 Screw the C mount adapter firmly into the video camera to be used 3 Place the C mount adapter with video camera on the original position and lightly tighten th...

Страница 7: ...aration to the transmitted light and reflected light ports Applicable UIS Mirror Units UIS fluorescence mirror units for BX BX2 series Attachable by removing the excitation filter frame and barrier filter UIS mirror unit frame U MF U MF2 Attach only a commercially available dichroic mirror and do not attach the excitation filter frame The fluorescence mirror unit for BX3 series can t be attached T...

Страница 8: ... into the threaded screw hole on the mount dovetail and slide out the mount dovetail to expose the mirror unit mount dovetail ƒ 4 Slightly loosen the mirror unit clamping screw using the Allen screw driver fit the dovetail on the mirror unit all the way into the mount dovetail ƒ and tighten the clamping screw firmly 5 Return the mount dovetail to the original position and remove the light path sel...

Страница 9: ...his adjustment 1 Looking into the diopter adjusted eyepieces bring the specimen into focus as accurately as possible 2 Using the Allen screwdriver lightly loosen the clamping screw of the C mount adapter for the transmitted light port Fig 4 3 Turn on the video camera and monitor 4 Slide the light path selector lever to select the light path for the trans mitted light port 5 Now the observation ima...

Страница 10: ...MEMO ...

Страница 11: ......

Страница 12: ...AX6301 04 8F Olympus Tower 446 Bongeunsa ro Gangnam gu Seoul Korea 135 509 Biological Industrial Distributed by Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan Manufactured by ...

Отзывы: