Fig. 9

Fig. 10

4-3  Observation Tube

1

Adjusting the Interpupillary Distance

(Fig. 9)

# Be sure to hold the binocular assembly @ with both hands to make

this adjustment.
To prevent damaging the mechanism, do not apply an excessive
force beyond the stop position.

While looking through the eyepieces, hold the left and right of the binocular
assembly @ and adjust the eyepieces by opening or closing them for
binocular vision until the left and right fields of view coincide completely.

2

Adjusting the Diopter
(Zoom Parfocal Adjustment)

(Fig. 10)

}Ensure that the eyepiece clamping knob @ is tightened before starting

the adjustment.
Adjusting the diopter specific to the two eyes of each observer allows the
parfocality to be ensured across zoom magnifications.

When Not Using the Eyepiece Micrometer Disk

1. Look into the right eyepiece and rotate its diopter adjustment ring ² so

that the peripheral part of the visual field looks sharply.

2. Place an easy-to-observe specimen on the stage plate.
3. Rotate the zooming knob ³ to a low magnification and, looking only into

the right eyepiece, focus the specimen using the coarse and fine focus
adjustment knobs.

4. Rotate the zooming knob ³ to the highest magnification and, looking

only into the right eyepiece, focus the specimen using the coarse and
fine focus adjustment knobs.

5. Rotate the zooming knob ³ to the lowest magnification, look only into

the left eyepiece, and focus the specimen by rotating the left diopter
adjustment ring | instead of the coarse and fine focus adjustment knobs.

When Using the Eyepiece Micrometer Disk

1. Look into the right eyepiece that includes the eyepiece micrometer disk,

and focus the micrometer disk by rotating the diopter adjustment ring ².

2. Place an easy-to-observe specimen on the stage plate.
3. Rotate the zooming knob ³ to the highest magnification and, looking

only into the right eyepiece, focus the specimen using the coarse and
fine focus adjustment knobs.
Ensure that both the eyepiece micrometer disk and specimen are focused
accurately.

4. Rotate the zooming knob ³ to the lowest magnification, , look only into

the left eyepiece, and focus the specimen by rotating the left diopter
adjustment ring | instead of the coarse and fine focus adjustment knobs.

}Note (or memorize) the diopter readings of the left and right eyepiece

scales so that they can be duplicated quickly in the next observation.

@

@

²

³

|

Содержание SZX16

Страница 1: ...his instruction manual is for the Olympus SZX16 Research High Class Stereo Microscope System To ensure the safety obtain optimum performance and to familiarize yourself full with the use of this micro...

Страница 2: ...This device complies with the requirements of directive 98 79 EC concerning in vitro diagnostic medical devices CE marking means the conformity to the directive...

Страница 3: ...Stop Function 3 Adjusting the Aperture Iris Diaphragm 4 Zoom Magnification Indication 5 2X Objective Correction Collar 4 3 Observation Tube 10 12 1 Adjusting the Interpupillary Distance 2 Adjusting t...

Страница 4: ...p Prevention Collar SZX R and Auxiliary Pillar SZH P400 SZH P600 22 9 2 Revolving Nosepiece SZX2 2RE16 23 25 9 3 BX Stage Adapter Type 1 SZX STAD1 26 28 9 4 BX Stage Adapter Type 2 SZX STAD2 29 30 9 5...

Страница 5: ...s not turn over Fig 1 2 To prevent the microscope body from turning over its left right pivot angle must be limited to 30 as shown in Fig 1 The tilt of the desktop surface should be no more than 5 3 S...

Страница 6: ...it hits a limiting mechanism the internal mechanism may be damaged 2 If the knobs on the left and right are rotated in opposite directions the internal mechanism will be damaged The rotation tension...

Страница 7: ...nctions or reduced performance 4 When not using the microscope keep it covered with the dust cover provided If a module containing a source of heat is mounted on the microscope wait until it has coole...

Страница 8: ...ocusing assembly SZX2 FO Coarse fine focusing assembly for heavy load SZX2 FOFH Motorized focusing assembly SZX2 FOA2 Intermediate attachments Coaxial vertical illuminator SZX2 ILLC16 Beam splitter SZ...

Страница 9: ...Click stop ON OFF screw P 8 Engages or disengages the click stop function for each zoom magnification Light path selector knob P 12 Zooming knob 0 7X to 11 5X Coarse focus adjustment knob tension adju...

Страница 10: ...y distance Page 10 3 Make diopter adjustment of the eyepieces Page 10 The adjustment procedure is variable depending on whether the eyepiece micrometer disk is used or not 4 Set the zooming knob to th...

Страница 11: ...ard to move because of the weight of option modules and or TV camera mounted on the microscope body it is recommended to use the focusing assembly for heavy load SZX2 FOFH The tension of the coarse fo...

Страница 12: ...ew fully by three turns from the ON position in the direction opposite to the arrow using the Allen screwdriver Do not rotate the screw too much or the cover may be damaged 3 Adjusting the Aperture Ir...

Страница 13: ...s water and plastic container When observing a specimen in a liquid or through a petri dish cover turn the collar to obtain the optimum contrast Correction is possible for aberration corresponding to...

Страница 14: ...he specimen using the coarse and fine focus adjustment knobs 4 Rotate the zooming knob to the highest magnification and looking only into the right eyepiece focus the specimen using the coarse and fin...

Страница 15: ...eyepiece 2 Clean the eyepiece micrometer disk 24 mm thickness 1 5 mm to remove dust and dirt then place the disk into the ring so that the side with reticule faces downward 3 Attach the ring with the...

Страница 16: ...Right Binocular 100 TV photo 100 Left Binocular 100 Always push or pull the light path selector knob all the way into a stop position Do not attempt to force the knob past the stop position Applying...

Страница 17: ...photomicrography areas when the WHN10X eyepieces with FN 22 are used 1 in CCD 2 3 in CCD 1 2 in CCD WHN10X FN 22 When the 0 5X TV adapter is used When the 1X TV adapter is used 2 Mounting the TV Adapt...

Страница 18: ...is visible in the field of view Dirt dust on the specimen 5 Details of observed image are solid Clean thoroughly Dirt dust on eyepiece The aperture iris diaphragm is stopped down excessively Open the...

Страница 19: ...stroke per turn 21 mm Fine focus adjustment knob stroke 80 mm Fine knob stroke per turn 0 77 mm Load 8 to 25 kg Load 2 7 to 15 kg Max load 10 kg 3 Observation tube SZX2 TR30 SZX2 TR30PT SZX2 TTR SZX2...

Страница 20: ...micrometer disk can be inserted Note The micrometer area outside the FN is invisible WHN10X H FN 22 diopter adjustment ring provided WHSZ15X H FN 16 diopter adjustment ring provided WHSZ20X H FN 12 5...

Страница 21: ...APO0 5XPF SDFPLAPO0 8X SDFPLAPO1XPF SDFPLAPO1 6XPF SDFPLAPO2XPFC 2 1X 104 8 3 15X 76 2 4 2X 59 5 6 3X 33 3 34 5X 6 4 51 75X 4 6 69X 3 6 103 5X 2 0 3 5X 62 9 5 25X 45 7 7X 35 7 10 5X 20 0 57 5X 3 8 86...

Страница 22: ...standard base and focusing assembly The Allen screwdriver can be stored in the pillar Do not use the Allen wrench for a purpose other than installing the pillar and the zoom microscope body Eyepieces...

Страница 23: ...ly clamping knob completely and while holding the focusing assembly with both hands insert the pillar into the mounting hole from below Fig 17 Insert slowly Do not apply excessive force 2 Lower the fo...

Страница 24: ...or an accident in case it is dropped Attach the cap to the objective tip Place the cap of the objective mount thread or a notebook etc on the stand to absorb the shock even when the objective drops Wh...

Страница 25: ...ge Plate Fig 23 Place the stage plate into the mounting hole on the base To remove press the stage plate at the edge nearest to the pillar with your fingertip The opposite end will rise from the base...

Страница 26: ...mination unit and SZX LGH1 light guide can be mounted When the specimen height is low or an objective with a short working distance is used the drop prevention collar does not have to be mounted if yo...

Страница 27: ...rnal View Nosepiece mount Hand guard Allen wrench for M3 screws Objective mounts Clamping screws M3 6 mm long x 4 2 Assembly Figs 27 to 29 1 Remove the observation tube from the zoom microscope body 2...

Страница 28: ...clamp the nosepiece mount while rotating it Fig 28 5 Install the focusing assembly on the pillar and install the observation tube in the original position 6 Attach two objectives onto their respective...

Страница 29: ...ng the left light path for photomicrography When the SZX2 LBS beam splitter is used the left light path can be used as the photo light path of optical axis aligned photomicrography 1 Rotate the desire...

Страница 30: ...Mountable Bases Restrictions Base Applicable Objectives Restrictions Standard base SZX2 ST SZX ST 0 5X to 2X None Transmitted illumination base SZX2 ILLK SZX ILLK Advanced illumination base SZX2 ILLB...

Страница 31: ...ting threaded holes x 2 Base Mounting the Polarizer SZX PO When simplified transmitted polarized light observation is required install the SZX PO polarizer on the SZX STAD1 BX stage adapter type 1 To...

Страница 32: ...d light observation 1 Loosen the two analyzer clamping knobs fit the analyzer into the objective tip and tighten the clamping knobs by positioning them on both sides 2 Without a specimen rotate the an...

Страница 33: ...er with the drop prevention collar The U SVL or S SVR BS stage for the BX series microscopes can also be mounted but the operability deteriorates The U SVLB and U SVRB cannot be used due to their long...

Страница 34: ...oblique illuminator remove the frosted filter and while observing through the eyepiece tilt the mirror gradually until optimum contrast can be obtained 3 Caution Do not project the image of the exter...

Страница 35: ......

Страница 36: ...ny 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A One Corporate Drive Orangeburg NY 10962 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilb...

Отзывы: