DMTA-20043-01EN [U8778624], Rev. B, May 2013
58 Chapter 6
Table 11 on page 58 details the data specifications of processing, TCG, and storage.
Display
Refresh rate
A-scan: 60 Hz; S-scan: 60 Hz
Envelope (echo-
dynamic mode)
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
On internal clock
1 Hz to 6 kHz
External pace
Yes
On encoder
On two axes: from 1 to 65536 steps
a. The interpolation is disabled for the TOFD channel, and when FFT
is enabled.
Table 11 Data specifications
Processing
Number of data points Up to 8192
Real-time averaging
PA: 2, 4, 8, 16
UT: 2, 4, 8, 16, 32, 64
Rectification
RF, full wave, half wave+, half wave–
Filtering
PA channel: 3 low-pass, 3 band-pass,
and 5 high-pass filters
UT channel: 3 low-pass, 6 band-pass,
and 3 high-pass filters (8 low-pass
filters when configured in TOFD)
Video filtering
Smoothing (adjusted to the probe
frequency range)
Programmable TCG
Number of points
16: One TCG (time-corrected gain)
curve per focal law
Range
PA: 40 dB per step of 0.1 dB
UT: 100 dB per step of 0.1 dB
Maximum slope
40 dB/10 ns
Table 10 Acquisition specifications
(continued)
Содержание OmniScan SX
Страница 6: ...DMTA 20043 01EN U8778624 Rev B May 2013 vi Table of Contents ...
Страница 8: ...DMTA 20043 01EN U8778624 Rev B May 2013 viii List of Abbreviations ...
Страница 24: ...DMTA 20043 01EN U8778624 Rev B May 2013 16 Important Information Please Read Before Use ...
Страница 26: ...DMTA 20043 01EN U8778624 Rev B May 2013 18 Introduction ...
Страница 58: ...DMTA 20043 01EN U8778624 Rev B May 2013 50 Chapter 4 ...
Страница 68: ...DMTA 20043 01EN U8778624 Rev B May 2013 60 Chapter 6 ...
Страница 80: ...DMTA 20043 01EN U8778624 Rev B May 2013 72 List of Figures ...
Страница 82: ...DMTA 20043 01EN U8778624 Rev B May 2013 74 List of Tables ...