DMTA-20043-01EN [U8778624], Rev. B, May 2013
56 Chapter 6
Table 8 on page 56 and Table 9 on page 56 detail the acoustic specifications of the
pulser, receiver, and beam formation, for both UT and PA modes.
When the UT channel is used in pulse-echo mode, pulses occur on both the
P
1 and
R
1
connectors. When pulse-echo mode is selected, Olympus recommends using only the
P
1 connector to connect the probe.
Table 8 Acoustic specifications — UT channel using UT connector
Pulser
Voltage
95 V, 175 V, and 340 V
Pulse width
Adjustable from 30 ns to 1000 ns;
resolution of 2.5 ns
Fall time
< 10 ns
Pulse shape
Negative square pulse
Output impedance
< 30
Ω
Receiver
Gain range
0 dB to 120 dB maximum input signal
34.5 Vp-p (full-screen height)
Input impedance
60
Ω
in pulse-echo mode
50
Ω
in pulse-receive mode
System bandwidth
0.25 MHz to 28 MHz (–3 dB)
Table 9 Acoustic specifications — PA channel
Pulser
Voltage
40 V, 80 V, and 115 V
Pulse width
Adjustable from 30 ns to 500 ns;
resolution of 2.5 ns
Fall time
< 10 ns
Pulse shape
Negative square pulse
Содержание OmniScan SX
Страница 6: ...DMTA 20043 01EN U8778624 Rev B May 2013 vi Table of Contents ...
Страница 8: ...DMTA 20043 01EN U8778624 Rev B May 2013 viii List of Abbreviations ...
Страница 24: ...DMTA 20043 01EN U8778624 Rev B May 2013 16 Important Information Please Read Before Use ...
Страница 26: ...DMTA 20043 01EN U8778624 Rev B May 2013 18 Introduction ...
Страница 58: ...DMTA 20043 01EN U8778624 Rev B May 2013 50 Chapter 4 ...
Страница 68: ...DMTA 20043 01EN U8778624 Rev B May 2013 60 Chapter 6 ...
Страница 80: ...DMTA 20043 01EN U8778624 Rev B May 2013 72 List of Figures ...
Страница 82: ...DMTA 20043 01EN U8778624 Rev B May 2013 74 List of Tables ...