7. Focus
1) Focusing with the OM-System Focusing Screen (Cross Hairs-clear type No. 1-12)
(1) Use the OM-System Dioptric Correction Lens or the Varimagni Finder (optional) so that
the double cross line within the field of view can be clearly recognized as two lines.
(2) Use the coarse and fine adjustment controls of the microscope to bring the specimen
in focus. Check again if both the double cross line and the specimen are equally sharp
in focus.
8. Determine the exposure.
1) Rotate the shutter ring or adjust light intensity by means of the light source or ND filters,
until the needle in the viewfinder comes to the center of index marks.
2) In most cases, it is advisable to rectify exposures in photomicrography, as follows:
(1)When the object to be photographed in the field of view is extremely small, set it for
overexposure.
(2)
When the most area in the field of view is dark and contrast of brightness is high,
adjust for under-exposure.