Chapter 3 Preparation and Inspection
29
BF TYPE 160 Series OPERATION MANUAL
1.
Inspect the sterile package for tears, inadequate sealing or water damage. If
the sterile package shows any irregularities, the sterile condition of the
instrument may have been compromised. Use a spare instead.
2.
Inspect the biopsy valve for cracks or other damage (see Figure 3.4).
Figure 3.4
Inspection of the mouthpiece (MA-651)
Do not use a mouthpiece that is damaged, deformed or
reveals other irregularities. Doing so may cause patient injury
and/or equipment damage.
Placing the mouthpiece in the patient’s mouth before the
procedure prevents the patient from biting and/or damaging
the endoscope’s insertion tube.
Housing
Slit