DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
224
c)
Fine-tune the
MIX
angle (on the
MIX
page of the
MAIN FILTER
menu) to
also obtain 40.0 degrees or very close to this value (see Figure 5-147 on
page 224).
Figure 5
‑
147 Fine
‑
tuning the MIX angle
2.
Press the FULL NEXT key (
) to activate the full-screen mode, and then scan the
five flat-bottom holes.
The signal should resemble the image shown in Figure 5-148 on page 224.
The list of all parameters is shown in Figure 5-149 on page 225.
Figure 5
‑
148 The scan of the flat
‑
bottom holes