DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
211
Figure 5
‑
127 Position 3 on the standard
22. Move the probe over notch B, and then, while holding the probe at this position,
adjust the
V GAIN 2
(E key) until the mix signal extends upward from the null
point 4 vertical divisions (see Figure 5-128 on page 211).
Figure 5
‑
128 Adjusting the V GAIN 2 value
23. Press the MAIN FILTER menu key (
) repeatedly until the
FREQ 1
page is
displayed.
24. Place the probe on position 1, press the A-LIFT NULL key (
), and then move
the probe back and forth over the next fastener on the left.
3
B