DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
167
•
PS5-AL indexing scanner (product discontinued)
•
Reflection probe for PS5: 200 kHz to 1 MHz, RD 0.250 (reflection differential
6.4 mm [0.25 in.]); P/N: 9217572 [U8616045]
•
Eddy current hole standard for demonstration purposes (not certified); P/N:
RSTD-10135 [U8863213]
To set the initial NORTEC 600 configuration
1.
Connect the probe to the rotating scanner (align the connector marks), and then
connect the scanner cable to both the rotating scanner and the PROBE connector
on the NORTEC 600.
2.
When prompted, select
CONTINUE
(A key) to accept the PowerLink
information.
3.
Press the ADV SETUP menu key (
) once, and then select
APPL SELECT
(A key) to open the application selection menu. Select
Indexing Scanner
with the
knob, then press
to accept (see Figure 5-63 on page 167).
Figure 5
‑
63 The Indexing Scanner application
To calibrate the signals
1.
Adjust the probe’s Z position (for indexing) so that it extends halfway down into
the calibration hole.
For more information on operating the PS5 indexing scanner, refer to the
scanner’s
User’s Manual
.
Содержание nortec 600
Страница 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Страница 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Страница 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Страница 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Страница 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Страница 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Страница 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Страница 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Страница 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Страница 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...