Using the Instrument
129
lift-off signal. However, make sure that you do not tilt the probe too much, since
this could cause damage (see Figure 5-12 on page 129).
Figure 5
‑
12 The bolt
‑
hole signal
3.
Set the signal angle according to one of the following two alternative methods.
Before proceeding, check which method is better suited to your needs:
— First alternative:
While holding the probe (still rotating) in the bad hole, press the ANGLE key
(
), and then adjust the angle with the knob in order to rotate the lift-off
noise to a horizontal position (see Figure 5-13 on page 129).
Figure 5
‑
13 Adjusting the lift
‑
off noise (first alternative)