DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument
123
Figure 5
‑
3 The automatic lift
‑
off function
5.
Press the FREEZE key (
).
You can now use both of your hands to complete the calibration.
6.
Press the ANGLE key (
). The
ANGLE
parameter defaults to the coarse
mode. Press the Enter key (
) in order to switch the
ANGLE
parameter to the
fine mode. Adjust the
ANGLE
value with the knob until the lift-off signal is as
close as possible to horizontal (see Figure 5-4 on page 123).
Figure 5
‑
4 The lift
‑
off signal as close as possible to horizontal