MV-1/MV-2 Endoscope Disinfectors
Frequently
Asked
Questions
09FQ0015 9/18
Two Corporate Center Drive, Melville, NY 11747
Copyright 2001, Olympus America Inc.
Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part
thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do
Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.
Reprocessing
See MV-1/MV-2 Endoscope Disinfectors User/Service Manual
See also document #09-WI-0241, MV-1/MV-2 Endoscope Disinfectors Preventive Maintenance.
In order to comply with the recommendations of your hospital’s infection control committee, and national and
local guidelines and policies, it may be necessary to disinfect or sterilize the MV-1/MV-2 and the water filtration
system. Instructions for this are provided in
paragraphs 8.18 and 8.19 of document # 09-WI-0239.
Also see the
MV-1/MV-2 Endoscope Disinfectors User/Service Manual.
All endoscopes must be precleaned before placement in the MV-1/MV-2. Follow the manufacturer’s instructions
and established professional guidelines (see
page 5 in the MV-1/MV-2 Endoscope Disinfectors User/Service
the MV-1/MV-2.
Caution:
Effective reprocessing can only be accomplished by complete manual cleaning of the endoscope
followed by exposure to a LCG.
Caution:
Duodenoscopes with elevator wire channels require additional cleaning and disinfection steps. It is
essential that the elevator wire channels of these endoscopes be manually cleaned, rinsed, and disinfected
before the endoscope is placed in the MV-1/MV-2. Refer to the duodenoscope’s reprocessing manual for
detailed instructions on cleaning, rinsing, and disinfecting the elevator wire channel.
Installation
1. Are all standard set components present?
#
list to be sure that all parts are present.
#
If parts are missing, contact Olympus.
See document #09-WI-0237 Product Specifications.
2. Is all ancillary equipment compatible?
#
list to be sure all ancillary equipment will work with the endoscope.
3. Has the equipment been properly disinfected or sterilized?
#
to disinfect or sterilize all equipment.