INSTRUCTIONS
SUCTION PUMP
KV--5
CAUTION
Federal (USA) law restricts this device to sale
by or on the order of a physician.
0086
Страница 1: ...INSTRUCTIONS SUCTION PUMP KV 5 CAUTION Federal USA law restricts this device to sale by or on the order of a physician 0086...
Страница 2: ...nearest Olympus office INTENDED USE The Olympus KV 5 suction pump is a simple reliable pump intended for aspiration use during flexible endoscopy and general medicalor surgicalsuction Thecompactsizeof...
Страница 3: ...1 STANDARD SETS FEATURES 1 2 OPERATING PRECAUTIONS 5 3 INSTRUCTIONS FOR USE 6 4 CLEANING CARE AND STORAGE 11 5 MAINTENANCE AND REPAIR 12 6 SPECIFICATIONS 14 7 TECHNICAL DESCRIPTION 15 8 ENVIRONMENTAL...
Страница 4: ...d set Single use suction system KV 5 suction pump Microbial Filter pack of 10 Power cable Suction jar 2 ltr K7503763 Lid for use with single use liner K7503761 Single use liner 2 ltr pk 25 K7503430 Pa...
Страница 5: ...g tube 2m pk 5 sterile Filter connecting tube 900mm pk 10 non sterile K7503657 Lid with float mechanism K7503760 Suction jar mounting kit for Olympus WM N60 D60 Olympus TC C1 NE K7503556 Suction jar m...
Страница 6: ...ecting tube 2m pk 50 sterile K7503432 Rest of World K7505595 Americas only Filter connecting tube 340mm pk 10 non sterile K7503486 For use with1 litre KV 5 mounted suction jar Filter connecting tube 9...
Страница 7: ...abels Label 1 CHANGE FILTER DAILY OR WHEN CONTAMINATED Label 2 SUCTION Label 3 MAXIMUM LEVEL OF VACUUM 85kPa HIGH VACUUM HIGH FLOW Label 4 RATING PLATE Label 5 CAUTION REMOVE TRANSIT SCREWS FROM BASE...
Страница 8: ...to be used must comply with local or national regulations D The KV 5 is not intended to be used when positioned on inclined surfaces as the suction jar fluid cut off mechanism may not function correct...
Страница 9: ...antity of water to dilute the jar contents D Single use suction liners used filters and suction tubing will contain potentially infectious material Dispose of carefully as dictated by local hospital p...
Страница 10: ...jar holder 1 litre suction jar 3 Remove the transit bung from the filter housing retain for transportation on the front of the unit and insert the microbial filter with the word Olympus facing outerm...
Страница 11: ...otective equipment and dispose of carefully as clinical waste for incineration Do NOT lift the suction jar by the lid otherwise the lid and jar may become separated and spillage of its contents may re...
Страница 12: ...with float 3 Referring to Figure 3 7 locate the suction jar into the suction jar holder located on the front panel of the unit NOTE The suction jar holder will only accommodate the 1 litre suction ja...
Страница 13: ...when the fluid reaches the 1 litre 2 litre mark NOTE When the fluid in the suction jar reaches the float the float will rise and seal the patient tube when capacity is reached preventing further sucti...
Страница 14: ...of these components will result 4 3 The suction jar may be sterilised by autoclaving up to 137 C max according to hospital policy or disinfected by cold fluid immersion CAUTION Do not use phenol or c...
Страница 15: ...ce Daily by user D Check that the filter connecting tube suction jar and lid are in good condition D Replace the microbial filter and or filter connecting tube with 90 connector if contaminated D Clea...
Страница 16: ...ism if used is not stuck in the closed position Check performance of the pump as per Section 5 1 Check the vacuum regulator is set to the correct position Check all connections and tubing for air leak...
Страница 17: ...a EEA In accordance with European Directive 2002 96 EC on Waste Electrical and Electronic Equipment this symbol indicates that the product must not be disposed of as unsorted municipal waste but shoul...
Страница 18: ...o the suction jar cap Vacuum gauge Range 0 to 100 kPa Accuracy better than 2 5 of full scale Olympus is continually developing its product range and reserves the right to alter the above specification...
Страница 19: ...atient tube Suction jar Vacuum pump Exhaust silencer Vacuum regulator Vacuum gauge Filter tube Exhaust Single use suction liner 8 ENVIRONMENTAL PROTECTION There are no known risks associated with disp...
Страница 20: ...01702 465677 Telephone 01702 616333 2951 Ishikawa cho Hachioji shi Tokyo 192 8507 Japan Fax 0426 46 2429 Telephone 0426 42 2111 OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS SINGAPORE PTE LTD OLYMPUS MOSCOW LI...